Page 22 of 28
Gamma Vacuum DIGITEL MPCq Users Manual
Document 900034, Rev B
9.1 Hex
Command
9.2 Description
9.3 Data
Field
9.4 Response
9.5 Data/Response
Description
2D
TSP TURN ON
Turns on TSP programmed mode.
R
Where:
R is TSP (1-2)
DF
TSP GET ACTIVE FILAMENT
Reads active TSP filament.
R
N
Where:
R is TSP (1-2)
N is the active filament.
30
TSP GET SUBLIMATION TARGET LEVEL
Reads sublimation target power level.
R
X, P
Where:
R is TSP (1-2)
X is number of amps/watts
P is ‘A’ for Amps or ‘W’ for Watts
31
TSP GET LOWER PRESSURE
Reads lower TSP pressure. See pressure
window option.
R
Z.Ze-ZZ
Where:
R is TSP (1-2)
Z.Ze-ZZ is lower pressure
EB
TSP SET LOWER PRESSURE
Sets upper TSP pressure. See pressure
window option.
R, Z.ZE-ZZ
Where:
R is TSP (1-2)
Z.Ze-ZZ is lower pressure
EA
TSP SET UPPER PRESSURE
Sets upper TSP pressure. See pressure
window option.
R, Z.ZE-ZZ
Where:
R is TSP (1-2)
Z.Ze-ZZ is the upper pressure
82
TSP GET UPPER PRESSURE
Reads upper TSP pressure. See pressure
window option.
R
Z.Ze-ZZ
Where:
R is TSP (1-2)
Z.Ze-ZZ is the upper pressure
72
TSP GET ONTIME
Reads TSP sublimation on time.
R
D
Where:
R is TSP (1-2)
D is the on-time (duration) value in seconds.
73
TSP GET PERIOD
Reads TSP cycle period.
R
D
Where:
R is TSP (1-2)
D is the cycle period value in minutes.
74
TSP GET RUNTIME POWER LEVEL
Reads current power level.
X,P
Where:
X is number of watts/amps. P is 'W' for Watts or
'A' for Amps
78
TSP SET SUBLIMATION TARGET LEVEL
Sets sublimation target power level.
R, X, M
Where:
R is TSP (1-2)
X is number of amps/watts
M is 'A' for Amps or ‘W’ for Watts
79
TSP SET PARAMETERS
Configures TSP parameters.
R, M, N,
X.XE-XX,
Y.YE-YY, D, P
Where:
R is TSP (1 or 2)
M is cycle period in minutes (time between fires)
N is number of cycles
X.XE-XX is high pressure limit
Y.YE-YY is low pressure limit
D is sublimation duration in seconds
P is pressure window (0 = on, 1 = off/ignore)
8B
TSP SET SUPPLY
Configures TSP supply.
R, S
Where:
R is TSP (1-2)
S is supply (0-2), 0 indicates no supply, pressure
window is off
8C
TSP GET SUPPLY
Reads TSP supply.
R
S
Where:
R is TSP (1-2)
S is supply (1-2)