Page 14 of 28
Gamma Vacuum DIGITEL MPCq Users Manual
Document 900034, Rev B
12. ERROR / STATUS CODES
Figure 21. Display Error/Status Codes Screen
12.1 Error Example ‘STANDBY, Error 02, Vacuum loss, Detected’
‘STANDBY’ - Indicates supply HV can be turned on. In comparison, ‘DISABLED’ indicates supply HV cannot be turned on.
Turning on supply HV will clear all outstanding errors not requiring user intervention.
For example, error 02 indicates vacuum loss detected while pump was running. User can turn on HV and start pump again, no other action is
required.
In comparison, error 20 indicates safeconn (HV interlock) is not satisfied. The state will read ‘DISABLED’. User cannot turn on HV until
safeconn is satisfied.
‘Error 02’ - See “Table 1. Error / Status Codes” for description of the errors.
‘Vacuum Loss’ - Short description of the error.
‘Detected’ - Indicates error status. Detected means error was detected at some time in the past, but not active anymore.
In comparison, ‘Active’ indicates error was detected and still active at this time. Addition user intervention is required to clear the error.
Table 1. Error / Status Codes
1.1 Error/
Status Code
1.2 Description
01
Too many cooldown cycles occurred during pump starting.
02
Vacuum loss detected. Output voltage dropped to less than 1.2kV while pump was running.
03
Short circuit detected during pump starting.
05
Excess power detected. Excess amount of power delivered to the pump for the configured pump size.
07
Excess pump start time. The output voltage did not reach 2kV within maximum pump starting time of 5 minutes.
12
Pump thermal runaway detected. Significant drop in voltage detected during pump starting.
20
Safeconn (HV Interlock) not satisfied. HV cannot run.
21
HV interlock not satisfied or HV switch is off. HV cannot run. See section on digital inputs for details.
22
Pump size set to 0L/s. HV cannot run.
23
Supply not calibrated. Current metering might not be accurate.
26
Supply over-temperature detected. HV cannot run.
NOTE:
If error code is not listed in the table please contact Gamma Vacuum for more information.