Gamma DIGITEL MPCq User Manual Download Page 3

Page 3 of 28

Gamma Vacuum DIGITEL MPCq Users Manual

Document 900034, Rev B

1.  GENERAL INFORMATION

The DIGITEL™ MPCq is an ion pump power supply controller.
It is designed to operate up to two independent ion pumps along 
with a titanium sublimation pump (TSP).
It can accommodate any ion pump size up to 1200 l/s.
The front panel display provides direct readout of the ion pump 
voltage, current, or pressure.
Individual model specification information is located on our website 
at: 

www.gammavacuum.com

.

WARNING: 

Do not use unauthorized parts.  
Such parts may compromise safety. Contact 
Gamma Vacuum with any questions.

NOTE: 

Read this entire manual and follow installation instructions. 
Failure to do so may cause injury and/or may void warranty.

2.  APPROVALS

The DIGITEL MPCq is shown to meet the intent of Directive 89/336/
EEC for Electromagnetic Compatibility and Low Voltage Directive 
73/23/EEC for product Safety. Compliance was demonstrated to 
the following specifications as listed in the Official Journal of the 
European Communities:

• 

IEC 61010-1

• 

EN 61010-1

• 

UL 61010-1

• 

CAN/CSA-C22.2 No. 61010-1

3.  WARNINGS

WARNING:

SHOCK HAZARD. CAN 
CAUSE INJURY OR DEATH. 
REMOVE POWER BEFORE 
SERVICING.

ALERTE:

RISQUE DE CHOC. PEUT 
CAUSER DES BLESSURES 
OU LA MORT. RETIRER LA 
SOURCE D’ALIMENTATION 
AVANT LE SERVICE.

警告:

点検・修理の際は 必
ず全ての電源を お
切りください。 使
用者が感電した り、
死亡、または 重傷
を負う危険が あり
ます。

安全警告:

。触

可 能

致受

或死亡。 

修前去掉

 源。

ADVERTENCIA: 

PELIGRO POR DESCARGA. PUEDE 
CAUSAR LESIÓN O INCLUSO LA 
MUERTE. RETIRE Y DESCONECTE 
LA FUENTE DE ALIMENTACION 
ELECTRICA, ANTES DE PROCEDER 
AL SERVICIO DE REPARACIÓN, 
MANTENIMIENTO O REVISION 
INTERNA.

ACHTUNG: 

GEFAHR ELEKTRISCHER 
SCHLÄGE. VERLETZUNGS- ODER 
LEBENSGEFAHR. TRENNEN 
SIE ALLE ELEKTRISCHEN 
ANSCHLÜSSE VON DER 
SPANNUNGSVERSORGUNG 
BEVOR SIE ARBEITEN AN DEM 
GERÄT AUSFÜHREN.

WARNING: 

HEAVY OBJECT. TO AVOID 
MUSCLE STRAIN OR BACK 
INJURY, USE LIFTING AIDS 
AND PROPER LIFTING 
TECHNIQUES WHEN 
REMOVING OR REPLACING.

ALERTE: 

OBJET LOURD. POUR 
EVITER UNE TENSION 
MUSCULAIRE OU UN MAL 
DE DOS, UTILISER DES 
AIDES ET DES TECHNIQUES 
DE LEVAGE APPROPRIEES 
POUR L’ENLEVEMENT OU LE 
DEPLACEMENT.

警告:

大変重い装置です。 
筋挫傷、背部損 
傷等を防ぐ為、移 
動や移設の際は適 
切な技術や補助機 
器等を使用してく 
ださい。

安全警告: 

重物。

避免肌肉 

或背部受

, 

当移

时请

 

使用起重

设备

以及 

适当的起重技

ADVERTENCIA: 

OBJETO PESADO. PARA EVITAR UN 
SOBRE-ESFUERZO MUSCULAR O 
DAÑO FÍSICO, UTILIZE LA AYUDA 
DE ELEVADORES Y TÉCNICAS 
APROPIADAS PARA EL MANEJO 
DE OBJETOS PESADOS, CUANDO 
LO TRANSPORTE, DESPLAZE O 
CONSIDERE REEMPLAZARLO.

ACHTUNG: 

SCHWERES OBJEKT. 
ZUR VERMEIDUNG VON 
MUSKELZERRUNGEN ODER 
RÜCKENSCHÄDEN BEIM 
TRANSPORT GEEIGNETE 
HUBVOR-RICHTUNGEN UND 
HEBETECHNIKEN VERWENDEN.

WARNING: 

READ AND UNDERSTAND 
OPERATOR’S MANUAL 
BEFORE USING THIS 
MACHINE. FAILURE TO 
FOLLOW OPERATING 
INSTRUCTIONS COULD 
RESULT IN INJURY OR 
DAMAGE TO EQUIPMENT.

ALERTE: 

LIRE ET COMPRENDRE LE 
MANUEL D’OPERATION 
AVANT D’UTILISER CETTE 
MACHINE . NE PAS SUIVRE 
LES INSTRUCTIONS 
D’OPERATION PEUT CAUSER 
DES BLESSURES OU DES 
DEGATS A L’EQUIPEMENT.

警告:

この装置を使用さ 
れる前に必ず取扱 
説明書を熟読し理 
解した上でご使用 
ください。取扱説 
明書の通り操作を 
しなかった場合、 
装置が損傷、破損 
することがありま 
す。

安全警告: 

在使用

台机器前, 

请务

阅读

并理解 

“操作

手册 

(指南)”。如果 
未能遵循操作步

 

明,将可能

致 

设备

坏。

ADVERTENCIA: 

LEA, ESTUDIE, Y ENTIENDA BIEN EL 
MANUAL DE OPERACION, ANTES 
DE USAR ESTA MAQUINARIA. 
UNA FALLA POR NO SEGUIR LAS 
INSTRUCCIONES OPERATIVAS, 
PUDIERA RESULTAR EN DAÑO O 
PERJUICO DEL EQUIPO.

ACHTUNG: 

LESEN UND VERSTEHEN SIE 
DIE BEDIENUNGSANLEITUNG 
BEVOR SIE DAS GERÄT IN BETRIEB 
NEHMEN. FEHLBEDIENUNGEN 
KÖNNEN ZU VERLETZUNGEN 
FÜHREN ODER DIE AUSRÜSTUNG 
BESCHÄDIGEN.

Summary of Contents for DIGITEL MPCq

Page 1: ...DIGITEL MPCq CONTROLLER USERS MANUAL Document 900034 Rev B ISO 9001 2015 Certified...

Page 2: ...able 7 Command CRC Checksum 19 Table 8 Response CRC Checksum 19 Table 9 Gamma Protocol Commands 20 Table 10 Command Packet Struct Over Ethernet Connection 24 FIGURES Figure 1 MPCq Front Panel 6 Figure...

Page 3: ...OCEDER AL SERVICIO DE REPARACI N MANTENIMIENTO O REVISION INTERNA ACHTUNG GEFAHR ELEKTRISCHER SCHL GE VERLETZUNGS ODER LEBENSGEFAHR TRENNEN SIE ALLE ELEKTRISCHEN ANSCHL SSE VON DER SPANNUNGSVERSORGUNG...

Page 4: ...nect the high voltage cable with the power on After turning the power off allow at least one minute before disconnecting electrical equipment 5 Do not operate the controller without a proper electrica...

Page 5: ...le 8 HV CONTROL ION PUMP OPERATION 8 1 Prior to using an ion pump Confirm following Rough pump down to 1x10 4 Torr or less the lower the better 1x10 6 Torr is recommended See Rough Pump user manual fo...

Page 6: ...11 4 Supply Setup on page 8 S Configured pump size in l s See 11 4 Supply Setup on page 8 10 FRONT PANEL Figure 1 MPCq Front Panel 10 1 Description Front panel containts Main power on off rocker type...

Page 7: ...tches home screen to TSP view Button is available only if Remote TSP Controller is connected to the MPCq Controller Menu Invokes main menu Exit Clears screen hides control buttons Turn On Off Turns HV...

Page 8: ...bration zeroing process 11 5 Display Setup Figure 8 Display Setup Screen Available setup Brigtness Level Configures screen brightness level 1 10 Screensaver Configures screen saver on off state Turnin...

Page 9: ...iply current log value in step 1 by 1 Step 3 Add the offset value 7 7 8 0 3 Volts The analog output pin will read 0 3V Logarithmic Pressure Example Pressure 1x10 9 Torr Offset 11 Output state Normal S...

Page 10: ...t digital output is asserted OFF Pressure When pressure value gets above this point digital output is deasserted HV Error Function Digital output is asserted when supply error is detected HV Output St...

Page 11: ...gister map is available on Gamma Vacuum website under download section 11 12 Profibus Setup Figure 15 Profibus Setup Screen For Profibus information reference Profibus user manual document 900033 avai...

Page 12: ...f TSP cannot continue because Pressure Window is set but supply is off TSP Armed Outside the pressure window TSP cannot continue because Pressure Window is set but pump pressure is outside the pressur...

Page 13: ...s between each sublimation cycle Advance Options Configure Invokes TSP advance options window TSP1 TSP2 Selects TSP for which above parameters are configured 11 17 TSP Pressure Window Setup Figure 20...

Page 14: ...ymore In comparison Active indicates error was detected and still active at this time Addition user intervention is required to clear the error Table 1 Error Status Codes 1 1 Error Status Code 1 2 Des...

Page 15: ...request 13 3 Connector J115 NET Profibus optional Standard Sub D 9 pin female connector Connection to Profibus interface For Profibus information reference Profibus user manual document 900033 availa...

Page 16: ...connected 28 Not connected 29 Reserved do not use ground 30 ANALOG OUT 1 output analog range 0 to 10VDC 31 Ground 32 ANALOG OUT 2 output analog range 0 to 10VDC 33 Ground 34 ANALOG OUT 3 output analo...

Page 17: ...e same serial port due to hardware loading limitations NOTE Must be supplied even when running RS 232 3 COMMAND CODE 2 ASCII hex characters Range 00 through FF See Table 9 Gamma Protocol Commands on p...

Page 18: ...ror Codes on page 18 NOTE If there is not an applicable error code a 00 will be transmitted 4 RESPONSE DATA As needed ACSII printable characters only Data field s are used to respond to commands reque...

Page 19: ...fy checksum for the response perform similar calculations Table 8 Response CRC Checksum 8 1 Characters 8 2 Value Decimal 8 3 Value Hex 0 48 0x30 1 49 0x31 space 32 0x20 O 79 0x4F K 75 0x4B space 32 0x...

Page 20: ...fied the current name is returned S TEXT TEXT Where S is supply 1 2 TEXT is string up to 15 characters FF SYS RESET Restarts the system N Where N is the reboot mode 0 Reboot 1 Reboot and start boot lo...

Page 21: ...Relay TTL output de energized Off Pressure must be at least 20 greater than On Pressure If not controller will automatically set Off Pressure to 20 greater than On Pressure 58 HV GET SET DIGITAL INPU...

Page 22: ...RE Reads upper TSP pressure See pressure window option R Z Ze ZZ Where R is TSP 1 2 Z Ze ZZ is the upper pressure 72 TSP GET ONTIME Reads TSP sublimation on time R D Where R is TSP 1 2 D is the on tim...

Page 23: ...e R is TSP 1 2 X is option Currently option 0 is only supported YY is a two digit status code 00 for TSP Initializing 01 for TSP Off 02 for TSP Armed Running 03 for TSP Armed Waiting pressure window s...

Page 24: ...if the first digit is a zero 4 DATA As needed ACSII printable characters only Data field s are for any commands that have a data value Not all commands require data field If command has more than one...

Page 25: ...er request SAFECONN Connector HV Interlock SMB with SHV 10 output connector or Mini Banana with Fisher output connector Display Controls 7 TFT WVGA 800x480 Color LCD 16 9 aspect ration Touchscreen Com...

Page 26: ...ma Vacuum free from defects in material and workmanship As Buyer s sole exclusive remedy under this warranty Gamma Vacuum agrees to either repair or replace at Gamma Vacuum s option and free of parts...

Page 27: ...on can be provided Original Purchase Order Number Gamma Vacuum Sales Order Number Product Order Number and or Product Description Product Serial Number All products received for repair or replacement...

Page 28: ...rimental to the operation of UHV products Hazardous substance means a chemical or substance or mixture of chemicals or substances which is regulated by the Federal Occupational Safety and Health Admin...

Reviews: