© Galvin Engineering Pty Ltd
Product Installation Guidelines
Version 1, 19 October 2022, Page
9
of
25
FUNCTION
PROCEDURE
FEEDBACK
SIGNAL
INFORMATION
DEFAULT
SETTING
RANGE
Via IQUA
Touch
First startup
After installation
1x
app. 4 sec. till
Remove hand/object from sensor
area
-.
-
wait
+
2x/sec., keep out of sensor area
wait till
1x
Confirmation, ready for operation
Temporary off
«Cleaning mode»
Activate
1x
app. 2 sec. till
1x
Extended function mode is active
2 min.
2x
a app. 0.5 sec
Pulsing during active function
Temporary off
Stop
1x
app. 2 sec.
1x
Manual stop
or automatic
1x
Automatic after preset time
Continuous run
Activate
1x
app. 2 sec. till
1x
Extended function mode is active
5 min.
0.5-20 min.
(in 0.5
steps) min.
1x
minimum 3 sec
.
(max. 5 sec.)
Water flows after releasing
Continuous run
Stop
1x
app. 2 sec.
Manual stop
or automatic
Automatic after preset time
line purge feature
activate /
deactivate
<< Automatic
flush on set
interval>>
(time of
waterflow: 3 min)
1x
app. 2 sec. till
1x
extended function mode is active
12h
Off
12 h
24 h
48 h
1x
app. 25 sec. till
1x - 4x
1x= off,
2x=12h,
3x= 24h,
4x=
48h ignore flashes after 5 sec. and 10
sec.
–
keep touched till
1x - 4x
after app. 25 sec., releasing defines
interval
Wait till
1x
Confirmation, ready for operation
restart electronics
1x
app. 2 sec. till
1x
extended function mode is active
2x
a app. 0,5 sec
pulsing during active function
1x
app. 5 sec. till
4x
1x/sec.
Release and wait till
+
1x/sec., keep out of sensor area
Wait till
1x
confirmation, ready for operation
8.0
COMMISSIONING
Initial Test for Correct Operation
The following must be checked to ensure the correct performance of the progressive thermostatic mixer:
−
the supply pressures and temperatures are within the range of operating pressures and temperatures for
the thermostatic valve.
−
the supply temperatures are within the range permitted for the unit.
Please record the testing done during the commissioning and record the temperature of the hot and cold-water
supplies and the temperature of the mixed water at the full hot setting.
A calibrated digital thermometer having rapid response time with maximum temperature hold will be required to
check and set the outlet mixed temperature of the mixer.
To Commission
To test the temperature, open the valve and allow the mixed heated water to flow for at least 30 to 60 seconds
so the temperature can stabilise before taking a temperature reading at the outlet with a digital thermometer. The
flow rate should be at least 4L/min.
The temper ture sh u d e t ken se t the m xer’s ut et nd f the ut et temper ture requ res adjustment,
then follow the temperature adjustment in the installation instructions.
XXX
XXX