Revision 5
56
August 2021
Figure 33: Integration Period Dialog Box
The current value of the integration time is displayed in the
Old Value
field. A new integration
period can be entered into the
New Value
field using the white numerical keys and the decimal
point key. Once the desired value has been entered into the New Value field, press ENTER,
then highlight the
Okay
button and press ENTER again. To discard changes, highlight the
Cancel
button and press ENTER.
After manually editing the integration period, check the indicated Peak Height. If the peak height
displays 32000, this indicates that the peak height is off scale, as the full scale for the
spectrometer reading is 32000 A/D counts. The integration period should be reduced until the
indicated peak height drops below 32000. If the indicated peak height is less than 24000, the
integration period should be increased. If the integration time is already at 1000 ms, this
indicates a problem with the analyzer’s optical system. Refer to Section 7.2 for more details on
troubleshooting analyzer problems.
4.6.1 Factory Reference
The
Factory Reference
is a zero gas transmission spectrum that was saved at the factory
at the time of analyzer factory calibration. When the
Show Factory Reference
checkbox is
checked by using the FIELD NEXT / FIELD PREV keys to highlight and then pressing
ENTER to check the box, the factory reference spectrum is shown superimposed on the
same axes as the currently displayed spectrum. The Spectrum panel with the Factory
Reference being displayed is shown in Figure 34.
Figure 34: Factory Reference
Summary of Contents for 943-TGXeNA
Page 130: ...Revision 5 130 August 2021 Figure 82 Oven Cabinet Door Removed ...
Page 131: ...Revision 5 131 August 2021 Figure 83 Control Cabinet Door Removed ...
Page 132: ...Revision 5 132 August 2021 Figure 84 Power Steam Air Signals Connection Details ...
Page 133: ...Revision 5 133 August 2021 Figure 85 AC Wiring Schematic ...
Page 134: ...Revision 5 134 August 2021 Figure 86 DC Signals and Wiring Diagram ...
Page 135: ...Revision 5 135 August 2021 Figure 87 Flow Diagram ...