Revision 5
118
August 2021
3. Remove the top O-ring. If it is not on the cell window, then it will be in the end cap O-
ring groove. Discard if there is any sign of damage or deformation. A fine tipped tool
may be required to remove the O-Ring from the groove in the cell end fitting. The cell
window may also stick to the cell end fitting when it is removed.
4. If the cell window did not come out with the cell end fitting, turn the measurement cell
over, holding your hand over the open end, and gently shake until the cell window
drops into your hand. Set the window aside on a clean surface. Remove the bottom O-
Ring (if it didn’t drop into your hand with the window). A fine tipped tool may be required.
Discard the O-ring if there is any sign of damage or deformation.
The cell window may stick to the inner O-Ring and be difficult to remove. If
it is, cap all but one port on the top of the measurement cell and blow into
the open port, holding a hand under the cell end to catch the cell window.
The air pressure should be sufficient to loosen the cell window and allow it
to be removed.
5. Repeat Steps 1-4 for the other cell window. If the other cell window sticks to the inner
O-Ring, it may be removed by passing a soft tipped tool through the measurement cell
and pushing on it gently from the inside until it comes loose.
6. Inspect both windows for damage and/or contaminants. If the windows are damaged
(chipped, cracked, discoloured, etc.) discard and replace them with new ones. If the
windows are not damaged, they should be cleaned with isopropyl alcohol, using a non-
abrasive lint free tissue such as KimWipes
®
or a cotton swab. Thoroughly rinse the
cleaned surface with pharmaceutical grade distilled water. They should be wiped dry
using a soft, lint free cloth or tissue.
Commercially available lens or window cleaning solutions or prepackaged
lens cleaning wipes ARE NOT to be used. These products may contain a
compound which leaves an UV absorbing coating on the cell window, which
could lead to continued High Integration Period warnings.
7. Inspect the measurement cell body. Clean with isopropyl alcohol, using a nonabrasive
lint free tissue such as KimWipes® or a cotton swab. Thoroughly rinse the cleaned
surface with pharmaceutical grade distilled water and dry completely before
reassembly.
8. Hold the measurement cell vertically and install the new bottom O-Ring in the top
window recess, ensuring that it is centred and sitting in the O-Ring seat.
9. Insert the cell window. Handle the window by its edges to prevent fingerprints on the
transparent surfaces of the window.
10. Place the top O-ring onto the window and replace the cell end fitting. The O-Ring will
self-align as the end fitting is tightened. Tighten the end fitting to hand-tight only.
Do not overtighten the cell end fittings. Overtightening these fittings could
cause damage or break the windows. Also ensure that there is no cross-
threading to avoid damaging the interior threads on the sample cell
block.The cell end fitting should screw on smoothly. DO NOT FORCE!
Summary of Contents for 943-TGXeNA
Page 130: ...Revision 5 130 August 2021 Figure 82 Oven Cabinet Door Removed ...
Page 131: ...Revision 5 131 August 2021 Figure 83 Control Cabinet Door Removed ...
Page 132: ...Revision 5 132 August 2021 Figure 84 Power Steam Air Signals Connection Details ...
Page 133: ...Revision 5 133 August 2021 Figure 85 AC Wiring Schematic ...
Page 134: ...Revision 5 134 August 2021 Figure 86 DC Signals and Wiring Diagram ...
Page 135: ...Revision 5 135 August 2021 Figure 87 Flow Diagram ...