Revision 5
143
August 2021
10. If there are any issues with the calibration discovered while performing the test
procedure in Step 9, repeat the calibration process in steps 4-8 and test again.
11. Repeat the calibration process steps 1-10 for each additional analog output that needs
to be calibrated.
12. When all analog outputs have been calibrated and tested, press
Permanent Save
to
save all changes permanently to the analyzer. This process will take approximately 30
seconds to complete.
Below the calibration and test fields for each analog are three read-only fields giving information
about the analog output. The
Description
field shows the analog output number (AO1 – AO4),
the
Unit
field shows the output measurement units (milliamps for all analog outputs), and the
Calculated Value
field shows the current output current, in milliamps, being output from that
analog output. The intensity of the green indicator LED on the IO board for each analog output
is directly proportional to the magnitude of the output current. That is to say, the LED will be
brighter when outputting 20 mA than when outputting 4 mA.
12.3.2 Testing the Digital Input
The IO board has a single digital input at terminal block P5 that is used for remote initiation of
a zero calibration cycle. This digital input can be tested using the procedure below.
1. Ensure that
Direct Board Control Enable
is selected on the
Manual Override
page.
2. Switch to the
Status
page.
3. Connect a switch with a 12-24VDC power supply to the digital input terminal block to
create an open circuit connected to the digital input terminal block. ‘
The digital input terminal is polarity insensitive.
4. When the switch is open and the circuit is not complete, the
Digital Input
indicator on
the
Status
page will be grey (inactive).
5. When the switch is closed and the circuit is complete so that 12-24VDC is being
received at P5, the
Digital Input
indicator on the
Status
page will be red (active).
12.3.3 Testing Digital Outputs
There are four digital (relay) outputs available at P4 of the IO board, as shown in Figure 92.
Summary of Contents for 943-TGXeNA
Page 130: ...Revision 5 130 August 2021 Figure 82 Oven Cabinet Door Removed ...
Page 131: ...Revision 5 131 August 2021 Figure 83 Control Cabinet Door Removed ...
Page 132: ...Revision 5 132 August 2021 Figure 84 Power Steam Air Signals Connection Details ...
Page 133: ...Revision 5 133 August 2021 Figure 85 AC Wiring Schematic ...
Page 134: ...Revision 5 134 August 2021 Figure 86 DC Signals and Wiring Diagram ...
Page 135: ...Revision 5 135 August 2021 Figure 87 Flow Diagram ...