6. SpO
2
Monitor (LX-7230KM)
40
■
Applying Universal Probe (UD-5C)
The UD-5C is suitable for long-term measurement, and can be attached to a
fingertip of an adult/ pediatric who is 40kg and above.
1.
Fix the probe with tape. Before attaching the probe to a patient, clean the
probe using a cloth moistened with sterilizing alcohol.
2.
Make sure that the two pads are aligned.
3.
Attach the pad with the LED to the base of the fingernail and the sensor
pad to the opposite side of the finger.
Light source (LED)
UD-5C LED
mark
Sensor
WARNING
Precautions for Use of Universal Probe (UD-5C)
Although UD-5C is suitable for long-term measurement, change the measuring
site every 8 hours (recommended wearing hours) to prevent rash or a low-
temperature burn. Be especially careful for continuous use on a patient with
peripheral circulatory disturbance by changing the measuring site frequently.
If the use of adhesive tape causes skin irritation, it should be discontinued.
When securing the probe with bandage tape, do not tighten it too hard as this
may cause edema or congestion.
When removing the tape, be careful not to damage the skin.
CAUTION
Precautions for Use of Universal Probe (UD-5C)
When removing the probe, do not pull on the cord, and remove the tape. Do not
apply excessive force to remove the tape. It may break the probe.
Cover the probe with a black cover if it is exposed to high-intensity light such as
direct light and surgical light to prevent measurement error.
Remove nail polish before taking measurements to enable stable
measurements.
It may not be possible to take measurements in case of an excessively high
hematocrit value. In such case, reattach the probe to a thinner finger.
After attaching the probe, make sure that there is no failure message and the
pulse wave level on the LCD is upper than 2. If not upper than level 2, it may
measure incorrectly. In such case, warm the finger to improve blood circulation.
Summary of Contents for LX-7230
Page 22: ...1 General Description 2 Blank Page ...
Page 28: ...3 Preparation 8 Blank Page ...
Page 36: ...5 Respiration Monitoring 16 Blank Page ...
Page 64: ...6 SpO2 Monitor LX 7230KM 44 Blank Page ...
Page 76: ...7 Measurement 56 Blank Page ...
Page 86: ...9 Other Setting Items 66 Blank Page ...
Page 88: ...10 Changing the Transmitter Channel and Group ID 68 Blank Page ...
Page 96: ...12 Cleaning and Disinfection 76 Blank Page ...
Page 111: ......