15. Specification
85
LX-7230KM
SpO
2
Measurement Range:
0 to 100%
Resolution: 1%
Measurement Accuracy:
Accuracy of measurement with SpO
2
probe
is as follows.
Measurement Spot
Measurement Accuracy
(±1SD)
Finger ±2%
Big toe or the root
of the big toe
±3%
(When SpO
2
is 70 to 100%. Less than 70% is not specified.)
Wavelength:
658 to 665nm (Red light), 860 to 925nm
(infrared light)
Power Loss on Probe:
2.0mW
Measurement Reply Time:
12 sec (moving average time)
Measurement Value Update Rate: Continuous mode 1 second
Intermittent mode 25 seconds
NOTE
The SpO
2
measurement accuracy is determined based on the values of
the root-mean-square (rms) difference between SpO
2
readings of the pulse
oximeter equipment and values of SaO
2
determined with a CO-oximeter,
by healthy adult volunteers. The pulse oximeter equipment measurements
are statistically distributed; ±2% measurement accuracy means that only
about two-thirds of pulse oximeter equipment measurements can be
expected to fall within ±2% of the value measured by a CO-oximeter.
Pulse Wave
LX-7230N
Pulse Rate Measurement Range:
20 to 250bpm
Measurement Accuracy
±3bpm
LX-7230KM
Pulse Rate Measurement Range:
30 to 230bpm
Measurement Accuracy
±2bpm (30 to 100bpm)
±2% (100 to 230bpm)
Measurement Response Time:
6 sec. (Moving Average Time)
Summary of Contents for LX-7230
Page 22: ...1 General Description 2 Blank Page ...
Page 28: ...3 Preparation 8 Blank Page ...
Page 36: ...5 Respiration Monitoring 16 Blank Page ...
Page 64: ...6 SpO2 Monitor LX 7230KM 44 Blank Page ...
Page 76: ...7 Measurement 56 Blank Page ...
Page 86: ...9 Other Setting Items 66 Blank Page ...
Page 88: ...10 Changing the Transmitter Channel and Group ID 68 Blank Page ...
Page 96: ...12 Cleaning and Disinfection 76 Blank Page ...
Page 111: ......