6. SpO
2
Monitoring (LX-7230N)
19
CAUTION
SpO
2
sensors for LX-7230N cannot be used on the LX-7230KM.
SpO
2
sensors are not waterproof. Keep away from liquids.
Do not pick up the device pulling the sensor or cable part. It may get
disconnected from the device and the device may be dropped.
A message is displayed when the SpO
2
sensor is disconnected from the
device.
A message is displayed when the device detects that the SpO
2
sensor is
disconnected from the patient. Properly attach the SpO
2
sensor to the
patient.
Do not reuse the single-use SpO
2
sensor. It may cause incorrect
measurements.
Read through the instruction of the SpO
2
sensor as well.
CAUTION
The accuracy of SpO
2
measurement may be influenced by abnormal
hemoglobin, such as carbon monoxide hemoglobin (COHb) and
methemoglobin (MetHb). It may be also affected by cardiogreen or
intravascular dyes.
In addition, the following case may affect the accuracy of SpO
2
and pulse
rate measurement.
Outside light (direct sunlight, fluorescent, light therapy equipment,
surgical light, infrared heat ramp, etc.)
Hypoperfusion
Excessive body motion (patient’s motion)
Pigment injected to the blood for testing
In case of measurement during receiving CPR treatment
Placement of SpO
2
sensor on limb with a blood-pressure cuff, arterial
catheter
External colorant such as nail polish
Abnormally low or high hemoglobin concentration
Venous pulse
Electrosurgery
Influence of electromagnetic waves from other electronics
High-intensity radio waves from cell phone
Summary of Contents for LX-7230
Page 22: ...1 General Description 2 Blank Page ...
Page 28: ...3 Preparation 8 Blank Page ...
Page 36: ...5 Respiration Monitoring 16 Blank Page ...
Page 64: ...6 SpO2 Monitor LX 7230KM 44 Blank Page ...
Page 76: ...7 Measurement 56 Blank Page ...
Page 86: ...9 Other Setting Items 66 Blank Page ...
Page 88: ...10 Changing the Transmitter Channel and Group ID 68 Blank Page ...
Page 96: ...12 Cleaning and Disinfection 76 Blank Page ...
Page 111: ......