xiii
EMC Guidance
This equipment complies with IEC 60601-1-2 (2007). However, if portable
transmitter or wireless LAN equipment is used extremely nearby, the
electromagnetic influence may largely exceed the compliance level and may
cause unexpected phenomenon such as noise interference on the waveform,
etc.
Therefore, this equipment should be used in a location specified by each
medical institution. If any unexpected noise interference on the waveform or
failure to the peripheral device occurs, stop using the equipment and follow the
instruction of the technician.
The following is the information relating to EMC (Electromagnetic
Compatibility).
(When using this equipment, verify that it is used within the environment
specified below.)
●
Compliance to the Electromagnetic Emissions
The LX-7230 is intended for use in the electromagnetic environment specified
below. The customer or the user of the LX-7230 should assure that it is used in
such an environment.
Emission Test
Compliance
Electromagnetic Environment –
Guidance
RF Emission
CISPR 11
Group 1
The LX-7230 uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF Emission
CISPR 11
Class A
Harmonic Emission
IEC 61000-3-2
N/A
Voltage Limit /
Flicker Emission
IEC 61000-3-3
N/A
This LX-7230 is suitable for use in all
establishments other than domestic and
those directly connected to the public low-
voltage power supply network that
supplies buildings used for domestic
purposes.
Summary of Contents for LX-7230
Page 22: ...1 General Description 2 Blank Page ...
Page 28: ...3 Preparation 8 Blank Page ...
Page 36: ...5 Respiration Monitoring 16 Blank Page ...
Page 64: ...6 SpO2 Monitor LX 7230KM 44 Blank Page ...
Page 76: ...7 Measurement 56 Blank Page ...
Page 86: ...9 Other Setting Items 66 Blank Page ...
Page 88: ...10 Changing the Transmitter Channel and Group ID 68 Blank Page ...
Page 96: ...12 Cleaning and Disinfection 76 Blank Page ...
Page 111: ......