11. Troubleshooting
74
SpO
2
Phenomenon Cause
Solution
The probe size is
improper.
Use a probe, which fit
properly.
The probe is
peeling off or is
affected by the
outside light due to
the poor condition
Attach the probe properly
following the instruction.
SpO
2
value is
unstable.
Transmitting and
measuring LEDs
sensor are dirty.
Clean both LED sensors
from dirt.
■
In Case of Dropping the LX-7230 into Water
In case of dropping the LX-7230 into water containing disinfectant, pick up the
LX-7230 quickly from it. Rinse it well with running water, and dry it thoroughly
with a soft cloth.
CAUTION
Do not use a drier. The LX-7230 shape may change or be broken.
When the LX-7230 is rinsed with running water, make sure to close the
battery compartment lid.