Chapter 12 Daily and Periodic Checks
12-24
Periodic Check List (1)
Carry out the periodic check according to "Periodic Check Procedure".
Management No.
Item Check
item
Procedure and measurement/
judgment criterion
Judgment
Note
(Repair required)
1
Exterior flaws, cracks,
deformation, rust
Check that flaws, cracks, deformation and rust
are not found on the main unit.
Good/ NG
2
Peeling and dirt on
nameplate and panel
Check that peeling or dirt is not found on the
nameplate or the panel.
Good/NG
Main uni
t
3 Key
Check that the keys are not damaged.
Good/NG
1 Power cable, patient cable
Check that flaws and damage are not found.
Good/NG
2
Limb electrode, chest
electrode
Check that dirt or rust is not attached and that
flaws and damage are not found.
Good/NG
3 Recording paper
Check that flaws and damage are not found.
Good/NG
V
is
ual ch
eck
Acces
so
ries
4 Operation
manual
Check that the manual is stored in the specified
storage place.
Good/NG
1 Key
Check that the key operation is smooth.
Good/NG
Ma
in
unit
2 Recorder
Check that the operation is smooth and that
abnormal sound is not heard.
Good/NG
1 Power cable, patient cable
Check the condition of the cables.
Good/NG
M
echani
cal
ch
ec
k
Acces
so
r
ies
2
Limb electrode, chest
electrode
Check that no problem is found with the
connection.
Good/NG
1 Status
Are all status normal?
Good/NG
2 Recorder printing test
Check that the printing paper is fed straight.
Check that both sides of the printing paper are
not creased.
Check that printing is made uniformly.
Check that there is no dead pixel.
Check that feeding speed is 5, 10, 12.5, 25 or
50mm/s (within ±3%)
Good/NG
3 Screen
display
test
Check that all the pixels are lighted and that
correct colors are displayed.
Good/NG
4 Key test
Do the keys respond correctly?
Good/NG
5 Sound test
Check that sound is emitted in correct volume.
Good/NG
6 Touch
panel
test
Check that the panel functions correctly at the
touch point.
Good/NG
7 Test pattern waveform
Are test pattern waveforms displayed normally?
Good/NG
Mainten
ance
Performance
8 RGB
test
Check that all the dots light and correct colors
are displayed.
Good/NG
1 Power
supply
Check that both AC power and battery power can
be turned on/off.
Good/NG
Others
2 Power
cable
Perform a continuity test using a tester and check
that the equipment is conducting.
Good/NG
Overall
judgment
Good/NG
Name of
equipment
Multi-function ECG
Installation site
Date of check
Model
Serial No.
FX-8322/FX-8322R
Inspected
by
Serial No.
Date of
purchase
Approved by
y
Copy this page for use.
Summary of Contents for CardiMax FX-8322
Page 4: ......
Page 16: ......
Page 17: ...Chapter 1 Overview C Ch ha ap pt te er r 1 1 Overview 1 Overview 1 1 ...
Page 18: ...Chapter 1 Overview ...
Page 20: ...Chapter 1 Overview 1 2 ...
Page 22: ...Chapter 2 Specifications ...
Page 28: ...Chapter 2 Specifications 2 6 ...
Page 30: ...Chapter 3 Names and Functions of Parts ...
Page 34: ...Chapter 4 Operational Description ...
Page 48: ...Chapter 5 Internal Connection Diagram ...
Page 50: ...Chapter 5 Internal Connection Diagram 5 2 ...
Page 51: ...Chapter 6 Spare Parts List C Ch ha ap pt te er r 6 6 Spare Parts List 6 Spare Parts List 6 1 ...
Page 52: ...Chapter 6 Spare Parts List ...
Page 56: ...Chapter 6 Spare Parts List 6 4 ...
Page 58: ...Chapter 7 Disassembly and Assembly Procedure ...
Page 75: ...Chapter 7 Disassembly and Assembly Procedure 7 17 7 17 Outside Drawing ...
Page 76: ...Chapter 7 Disassembly and Assembly Procedure 7 18 ...
Page 78: ...Chapter 8 Software Update Procedure ...
Page 81: ...Chapter 9 Caution Labels C Ch ha ap pt te er r 9 9 Caution Labels 9 Caution Labels 9 1 ...
Page 82: ...Chapter 9 Caution Labels ...
Page 84: ...Chapter 9 Caution Labels 9 2 ...
Page 86: ...Chapter 10 Troubleshooting ...
Page 98: ...Chapter 11 Maintenance ...
Page 114: ...Chapter 12 Daily and Periodic Checks ...
Page 140: ......