Chapter 12 Daily and Periodic Checks
12-7
12-3-1-2 Earth Leakage Current
Check the earth leakage current (current that flows to protective earth resistance).
z
Acceptance standards
Normal condition
0.5mA or less
Single fault condition
1.0mA or less
12-3-1-3 Enclosure Leakage Current
Check the enclosure leakage current (current that flows from the enclosure to the earth terminal of the power
outlet).
z
Acceptance standards
Normal condition
0.1mA or less
Single fault condition
0.5mA or less
12-3-1-4 Patient Leakage Current I
Check the patient leakage current I (a current that flows from the patient attachment to the protective earth
terminal).
z
Acceptance standards
Normal condition
0.01mA or less
Single fault condition
0.05mA or less
12-3-1-5 Patient Leakage Current III
Check the patient leakage current III (a current that flows from the external device to the patient attachment).
z
Acceptance standards
Single fault condition
0.05mA or less
12-3-1-6 Patient Measurement Current
Check the patient measurement current.
z
Acceptance standards
Normal condition
0.01mA or less
Single fault condition
0.05mA or less
Summary of Contents for CardiMax FX-8322
Page 4: ......
Page 16: ......
Page 17: ...Chapter 1 Overview C Ch ha ap pt te er r 1 1 Overview 1 Overview 1 1 ...
Page 18: ...Chapter 1 Overview ...
Page 20: ...Chapter 1 Overview 1 2 ...
Page 22: ...Chapter 2 Specifications ...
Page 28: ...Chapter 2 Specifications 2 6 ...
Page 30: ...Chapter 3 Names and Functions of Parts ...
Page 34: ...Chapter 4 Operational Description ...
Page 48: ...Chapter 5 Internal Connection Diagram ...
Page 50: ...Chapter 5 Internal Connection Diagram 5 2 ...
Page 51: ...Chapter 6 Spare Parts List C Ch ha ap pt te er r 6 6 Spare Parts List 6 Spare Parts List 6 1 ...
Page 52: ...Chapter 6 Spare Parts List ...
Page 56: ...Chapter 6 Spare Parts List 6 4 ...
Page 58: ...Chapter 7 Disassembly and Assembly Procedure ...
Page 75: ...Chapter 7 Disassembly and Assembly Procedure 7 17 7 17 Outside Drawing ...
Page 76: ...Chapter 7 Disassembly and Assembly Procedure 7 18 ...
Page 78: ...Chapter 8 Software Update Procedure ...
Page 81: ...Chapter 9 Caution Labels C Ch ha ap pt te er r 9 9 Caution Labels 9 Caution Labels 9 1 ...
Page 82: ...Chapter 9 Caution Labels ...
Page 84: ...Chapter 9 Caution Labels 9 2 ...
Page 86: ...Chapter 10 Troubleshooting ...
Page 98: ...Chapter 11 Maintenance ...
Page 114: ...Chapter 12 Daily and Periodic Checks ...
Page 140: ......