IP Handling
B-3
1
2
3
A
B
C
Y
Z
010-021-20 09.2001
2.3
Cleaning
(1) Materials required
Lint free non-woven cotton
Gauze (100% cotton),
Lens cleaner such as Toraysee
®
(2) Method
1
With a dry cloth of materials listed above, wipe off the IP surface as illustrated below (A
or B).
A
B
2
For stains which fail to come off by cleaning with dry cleaning cloths, dampen the
above cloth with ethanol anhydride (JIS standards or special standards) before clean-
ing. (Do not use another cleaning agent.)
When the IP surface has been cleaned using ethanol anhydride-dampened cloths, slightly
wipe it off with dry cloths listed above. Make sure the surface of the IP is completely dry
before use.
Z
Frequent cleaning with ethanol anhydride can cause yellowing of the IP edges.
(3) Precautions
1
Be sure not to damage the IP.
2
Never use eraser or any other solvents.
3
Do not scratch the IP with your nails.
Z
Do not apply light to the IP exposed but not yet processed.
Summary of Contents for FCR XG-1
Page 6: ...vi 010 021 20 09 2001...
Page 7: ...1 1 010 021 20 09 2001 1 Chapter 1 Introduction...
Page 17: ...2 1 010 021 20 09 2001 2 Chapter 2 Operations...
Page 37: ...3 1 010 021 20 09 2001 3 Chapter 3 When an Error Occurs...
Page 39: ...A 1 010 021 20 09 2001 A Appendix A Specifications...
Page 46: ...Specifications A 8 010 021 20 09 2001...
Page 47: ...B 1 010 021 20 09 2001 B Appendix B IP Handling...
Page 50: ...IP Handling B 4 010 021 20 09 2001...
Page 51: ...C 1 010 021 20 09 2001 C Appendix C Film Annotation...
Page 54: ...Film Annotation C 4 010 021 20 09 2001...
Page 91: ...Z 1 010 021 20 09 2001 Z Appendix Z Precautions for Exposure...
Page 100: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...