Specifications
A-3
1
2
3
A
B
C
Y
Z
010-021-20 09.2001
Applicable when X-ray dose is the same as chest exposure done with Hi-screen B-2/New
RX for ST.
(When the maximum dose applied to the IP exceeds 25mR for ST, the cycle time will
increase in accordance with the increase in dose. The above times are for processing one
cassette.)
(3) Reading gray scale
12 bits/pixel
Output gray scale
10 bits/pixel
(4) Reading Specification
IP auto feed/load mechanism cycle time
14
×
17” (35
×
43cm)
Approx. 64sec.
14
×
14” (35
×
35cm)
Approx. 56 sec.
10
×
12”
Approx. 52 sec.
8
×
10”
Approx. 42 sec.
24
×
30cm (ST)
Approx. 52 sec.
18
×
24cm (ST)
Approx. 40 sec.
2 Image Reading
(1) Available IP types and sizes
1. Inch settings
VI
(standard type):
14"
×
17", 14"
×
14",
10"
×
12", 8"
×
10"
2. Metric settings
VI
(standard type):
35
×
43cm, 35
×
35cm,
24
×
30cm, 18
×
24cm
Z
HR-V18
×
24cm IPs can only be used with metric settings, with 18
×
24M cassettes.
(2) IP auto feed/load mechanism cycle time
18
×
24cm
10
1770
×
2370
10
1770
×
2370
24
×
30cm
6.7
1576
×
1976
10
2364
×
2964
8"
×
10"
10
2000
×
2510
10
2000
×
2510
Reading Size
Standard
Sampling Rate
Pixel-density (pixels / mm)
Number of Pixels
High
Sampling Rate
Pixel-density (pixels / mm)
Number of Pixels
10"
×
12"
6.7
1670
×
2010
10
2505
×
3015
14"
×
14"
5
1760
×
1760
10
3520
×
3520
14"
×
17"
5
1760
×
2140
10
3520
×
4280
Summary of Contents for FCR XG-1
Page 6: ...vi 010 021 20 09 2001...
Page 7: ...1 1 010 021 20 09 2001 1 Chapter 1 Introduction...
Page 17: ...2 1 010 021 20 09 2001 2 Chapter 2 Operations...
Page 37: ...3 1 010 021 20 09 2001 3 Chapter 3 When an Error Occurs...
Page 39: ...A 1 010 021 20 09 2001 A Appendix A Specifications...
Page 46: ...Specifications A 8 010 021 20 09 2001...
Page 47: ...B 1 010 021 20 09 2001 B Appendix B IP Handling...
Page 50: ...IP Handling B 4 010 021 20 09 2001...
Page 51: ...C 1 010 021 20 09 2001 C Appendix C Film Annotation...
Page 54: ...Film Annotation C 4 010 021 20 09 2001...
Page 91: ...Z 1 010 021 20 09 2001 Z Appendix Z Precautions for Exposure...
Page 100: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...