Operations
2-18
010-021-20 09.2001
2.3
Conducting Exposure Using a Cassette (IP) Left Unused for 8 Hours or More
[SECONDARY ERASURE]
When conducting exposure with a cassette (IP) that has not been used for 8 hours or more,
the cassette (IP) should be subjected to "Secondary Erasure Processing" for erasing such
excess energy as natural radioactive rays accumulated on the IP. Operation procedures fol-
low.
1
Press the
Erasure Selection Switch
once to select "Secondary Erasure Processing."
Pressing this button once will select
"Secondary Erasure Processing."
will be displayed in the Reading Unit status display
area of the CR Console.
2
"Secondary Erasure Processing" has thus been specified for the cassette to be inserted.
Insert the cassette in the cassette setting unit. The IP contained in the cassette will be
subjected to "Secondary Erasure Processing."
As the IP returned to the cassette will have been undergone "Secondary Erasure
Processing," it can be now used for exposure.
Z
When "Secondary Erasure Processing" has been specified, the setting (display)
will be returned to default status ("Reading") at the stage the cassette is inserted.
The setting is thus valid one time only.
Cassette Insertion Lamp
must be lit.
Summary of Contents for FCR XG-1
Page 6: ...vi 010 021 20 09 2001...
Page 7: ...1 1 010 021 20 09 2001 1 Chapter 1 Introduction...
Page 17: ...2 1 010 021 20 09 2001 2 Chapter 2 Operations...
Page 37: ...3 1 010 021 20 09 2001 3 Chapter 3 When an Error Occurs...
Page 39: ...A 1 010 021 20 09 2001 A Appendix A Specifications...
Page 46: ...Specifications A 8 010 021 20 09 2001...
Page 47: ...B 1 010 021 20 09 2001 B Appendix B IP Handling...
Page 50: ...IP Handling B 4 010 021 20 09 2001...
Page 51: ...C 1 010 021 20 09 2001 C Appendix C Film Annotation...
Page 54: ...Film Annotation C 4 010 021 20 09 2001...
Page 91: ...Z 1 010 021 20 09 2001 Z Appendix Z Precautions for Exposure...
Page 100: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...