202B1263075
(4) Switch on the power to the processor and turn on the lamp.
(5)
Deflate the balloon on the distal end of endoscope and the
balloon on the Over-tube.
Apply clean lubricant (Xylocaine jelly or the like) to the
insertion portion as required.
[Note]
Do not apply Xylocaine spray, olive oil or the like directly
to the insertion portion.
(6) Insert the distal end of the endoscope from the oral cavity
to the pharynx, while observing the process.
Adjust the brightness with the level button on the light
source.
(7) Feed air to inflate the balloon on the distal end of the
endoscope, and stabilize it within the body cavity (intestinal
tract).
Chapter 4 Method of Use
80