2.5.2 Cleaning Adapter
The cleaning adapter (CA-610) is comprised of the injection tube (WA-007) and channel
plug (WA-010).
<Injection Tube (WA-007)>
Instrument/Suction Channel Connector
The syringe is attached to this connector to
feed air or fluid to the suction channel of the
endoscope.
Syringe
The syringe is attached to the
cleaning adapter to feed air or
detergent solution.
Air/Water Channel Connector
The syringe is attached to this connector to
feed air or fluid to the air/water channel of
the endoscope.
Tube for Instrument/Suction Channel
Connects to the suction channel.
Air Supply Tube Port
Connects to the air supply
tube of the LG connector
on the endoscope.
Tank Receiving Cap
Connects to the endoscope's water feed connector.
Tube for Air/Water Channel
Connects to the air/water channel.
Weight
Allows the syringe to feed detergent solution
while this weight is submerged under the detergent
solution, and to feed air while this weight is in the air.
Display Plate
Chapter 2 Product Overview
38
202B1263075
EI580BT_E2-10_202B1263075.indb 38
2015/08/12 16:20:35