4.6 Endoscope Withdrawal
(1) When a procedure is over, discharge any excessive air
from the body cavity.
(2) Unlock the up-down and left-right knobs.
(3) Operate the forceps stand lever to lower the forceps stand.
(4)
If the balloon is attached, deflate it completely.
[Note]
If the balloon does not deflate while depressing the suction
valve fully, remove clogs in the balloon channel using the
cleaning brush (WB2221FW2).
Reprocessing Manual
“4.9.4 Brushing the Balloon Channel”
(5) Straighten the bending section by operating the angulation
knobs.