Electromagnetic immunity compliance information and guidance
Immunity test
EN 60601-1-2
Test level
Compliance
level
Guidance
Electrostatic discharge (ESD)
EN 61000-4-2
± 6kV: contact
± 8kV: air
Same as left
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient/burst
EN 61000-4-4
± 2kV: for power supply lines
± 1kV: for input/output lines
Same as left
Main power quality should be that of a
typical commercial or hospital.
Surge
EN 61000-4-5
± 1kV: Line to line
± 2kV: Line to ground
Same as left
Main power quality should be that of a
typical commercial or hospital.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
EN 61000-4-11
< 11.5V (> 218.5V dip)
For 0.5 cycle
92V (138V dip)
For 5 cycle
161V (69V dip)
For 25 cycle
< 11.5V (> 218.5V dip)
For 5 sec
Same as left
Main power quality should be that of a
typical commercial or hospital. If the
user of this product requires continued
operation during power mains
interruptions, it is recommended
that this product is powered from an
uninterruptible power supply or battery.
Power frequency (50/60 Hz)
magnetic field
EN 61000-4-8
3 A/m
Same as left
It is recommended to use this product
by maintaining enough distance from
any equipment that operates with high
current.