8 Mode Settings and Functions
8-24
8
897N120152H FDC NX10N Instruction Manual 04.2022
(
ȣ
g/dl)
500
500
10
0
0
Y
X
Y=X
Y=0.98X + 5
(
ȣ
g/dl)
Measurement using
your instruments
FDC
reference
(
ȣ
g/dl)
(
ȣ
g/dl)
Measurement using
your instruments
FDC
reference
500
500
10
0
0
Y
X
Y=X
Y=1.1X - 5
(4)
Measurement range after inputting coefficients
(a, b)
Depending on the values input for (a, b), the
apparent limits of the measurement range will
shift as follows.
If a = 1 and b = 0 (Y = X), the measurement range is
10-500 μg/dl. But by inputting regression coefficients
as shown, the lower limit changes to A and the upper
limit to B.
If the regression formula is Y = 0.98X + 5:
A = 5 μg/dl
B = 505 μg/dl
If the regression formula is Y = 1.1X -5:
A = 14 μg/dl
B = 459 μg/dl
Summary of Contents for DRI-CHEM NX10N
Page 2: ......
Page 4: ......
Page 8: ...897N120152H FDC NX10N Instruction Manual 04 2022 iv Contents...
Page 18: ...1 Safe Usage and Handling Precautions 1 10 1 897N120152H FDC NX10N Instruction Manual 04 2022...
Page 24: ...3 Principles of Operation 3 2 3 897N120152H FDC NX10N Instruction Manual 04 2022...
Page 46: ...6 Quality Control 6 2 6 897N120152H FDC NX10N Instruction Manual 04 2022...
Page 54: ...7 Troubleshooting 7 8 7 897N120152H FDC NX10N Instruction Manual 04 2022...
Page 92: ...8 Mode Settings and Functions 8 38 8 897N120152H FDC NX10N Instruction Manual 04 2022...
Page 94: ...9 Other Functions 9 2 9 897N120152H FDC NX10N Instruction Manual 04 2022...
Page 98: ...10 Specifications Consumables 10 4 10 897N120152H FDC NX10N Instruction Manual 04 2022...
Page 100: ...11 Glossary 11 2 11 897N120152H FDC NX10N Instruction Manual 04 2022...
Page 101: ......