7 Troubleshooting
7-3
7
897N120152H FDC NX10N Instruction Manual 04.2022
7.2 Troubleshooting
NOTE:
- To stop beeping when an error occur, touch the
C (clear)
key.
- When an error occurs, the analyzer may terminate the measurement. When rerunning the sample, set the sample
ID and a slide, and spot the sample again from the beginning to restart the measurement.
7.2.1 Startup Errors
(1) When the analyzer does not start after the power switch is depressed to the [ | ] side:
Check that the power cable and AC adapter are properly connected and then turn on the analyzer again.
If the analyzer still does not start, please contact your technical support (dealer) for assistance.
7.2.2 Transfer errors
Error code
Error message
Error status
Solvent
E0200
Feed motor ERR
Sensor ON failure
When these errors occur, the
disposal box may be full of
slides, or a foreign object
may remain on the slide
transfer part.
Turn the power switch off, and
perform the following:
• Check whether there is any foreign
object at the photometer, slide
transfer part and transfer frame.
• Clean the photometer, slide transfer
part and transfer frame.
• Dispose of slides in the disposal
box. (Refer to P5-3 - 5-5)
If the error still occurs, please contact
your technical support (dealer) for
assistance.
E0201
Feed motor ERR
Sensor OFF failure
E0202
Feed motor ERR(Test)
Sensor ON failure
E0203
Feed motor ERR(Test)
Sensor OFF failure
E0204
Feed motor ERR(Eject)
Sensor ON failure
E0205
Feed motor ERR(Eject)
Sensor OFF failure
E0210
Feed motor ERR
Faulty controller
Contact your dealer
7.2.3 Photometer errors
Error code
Error message
Error status
Solvent
E0300
W-plate out of range
Clean photometer
Xn:****[digit]
X1:****[digit]
The circuit board has a
malfunction (abnormal
reference white plate
reading).
Turn the power switch off, and
perform the following:
• Clean the reference white plate and
the reference black plate.
• Clean the photometer.
(Refer to P5-3)
If the error still occurs, please contact
your technical support (dealer) for
assistance.
E0301
Abnormal BLK plate
Clean photometer
Xn:****[digit]
The circuit board has a
malfunction (abnormal
reference black plate
reading).
E0302
Dirty BLK plate
Clean photometer
***[digit]
The light source lamp unit,
reference black plate or
photometer head may be
dirty.
E0303
Light too intense
Clean photometer
****[digit]
The circuit board has a
malfunction (abnormal
reference white plate
reading).
E0304
Low light intensity
Clean photometer
****[digit]
E0305
Low light intensity
Clean photometer
****[digit]
Summary of Contents for DRI-CHEM NX10N
Page 2: ......
Page 4: ......
Page 8: ...897N120152H FDC NX10N Instruction Manual 04 2022 iv Contents...
Page 18: ...1 Safe Usage and Handling Precautions 1 10 1 897N120152H FDC NX10N Instruction Manual 04 2022...
Page 24: ...3 Principles of Operation 3 2 3 897N120152H FDC NX10N Instruction Manual 04 2022...
Page 46: ...6 Quality Control 6 2 6 897N120152H FDC NX10N Instruction Manual 04 2022...
Page 54: ...7 Troubleshooting 7 8 7 897N120152H FDC NX10N Instruction Manual 04 2022...
Page 92: ...8 Mode Settings and Functions 8 38 8 897N120152H FDC NX10N Instruction Manual 04 2022...
Page 94: ...9 Other Functions 9 2 9 897N120152H FDC NX10N Instruction Manual 04 2022...
Page 98: ...10 Specifications Consumables 10 4 10 897N120152H FDC NX10N Instruction Manual 04 2022...
Page 100: ...11 Glossary 11 2 11 897N120152H FDC NX10N Instruction Manual 04 2022...
Page 101: ......