NOTICE!
Place the Cross beater mill on a flat, stable surface. It
may be screwed to this or a base plate.
NOTICE!
Never operate PULVERISETTE 16 while it is standing on
the transport pallet!
The cross beater mill is mounted on an underframe. Remove the screws
and set the mill down on the required installation site with the help of
the ring bolt.
4.4 Ambient conditions
WARNING!
Mains voltage!
– The device may only be operated indoors.
– The surrounding air may not carry any electrically
conductive dust.
– Maximum relative humidity 80% for temperatures
up to 31°C, linearly decreasing down to 50% rela-
tive humidity at 40°C.
n
The room temperature has to stay between 5 - 40°C.
n
Altitudes up to 2000 m
n
Degree of pollution 2 according to IEC 664.
4.5 Setting up on a laboratory bench
You can mount the cross beater mill on a laboratory bench:
n
Measure the thickness of the laboratory bench.
n
Use screws with a maximum
∅
8 mm and a length equal to the labo-
ratory bench thi 25 mm.
n
Place the cross beater mill together with the collecting vessel onto
the bench.
n
Mark bore holes on the bench for drilling using the existing bore
holes of the U-beams beneath the cross beater mill.
The distance of the bore holes to the edge of the bench should be
chosen such that the collecting vessel and the filter hose can be readily
mounted and dismounted.
Installation
- 17 -