7 Technical data
22
KaVo Dental GmbH
214760 rev 1
DENTAL CARE UNIT MOUNT MODEL DEVICE
Horizontal arm length
330 mm (from axl to axl),
418 mm (total length)
Generator module length
500 mm
Horizontal arm shaft diameter
32 mm
Main cable length
5230 mm
(measured from generator
module to the scissors arm)
TUBE HEAD ASSEMBLY
Tube head assembly type
THA-I
Tube type
Toshiba D-0711SB or Kailong KL
21 SB or equivalent
Max. tube voltage
60 or 70 kV
Max. tube current
7 mA
Max. electric output
490 W (70kV; 7mA)
Reference axis
Runs axially with the cone
Target angle
16 degrees
Focal spot
0.7 mm
(according to IEC 60336/2005)
Nominal anode input power
940 W
Max. symmetrical radiation
field
60 mm at a 200 mm focal length
Total filtration
2,0 mm Al (70 kV)
Inherent filtration
1,5 mm Al (70 kV)
Fixed additional filtration
1,0 mm Al (70 kV)
Max. anode heat content
7 kJ
Maximum X-ray tube assembly
heat content
140 kJ
Maximum continuous heat
dissipation of the X-ray tube
head assembly
19 W
ELECTRICAL CONNECTIONS
Nominal mains voltage
115 VAC +/- 10%
230 VAC +/- 10%
Input power frequency
60 Hz
50 Hz
Nominal current
7.63A
3.55A
Mains fuse, slow blow
6.25AT
Apparent resistance of supply
mains
0.68
Power consumption
816 VA / 230 VAC
877 VA / 115 VAC
US/Canada mains connector
type
115 V / NEMA 6-15P or similar
Power supply cords type
H05VV5-F / AWG 14 (UL 2587)
Summary of Contents for 214760
Page 1: ...FOCUS Intraoral X ray User Manual ENGLISH 214760 rev 1 0 805 4936...
Page 2: ......
Page 4: ......
Page 6: ...ii KaVo Dental GmbH 214760 rev 1...
Page 21: ...2 Unit description 214760 rev 1 KaVo Dental GmbH 5 Exposure time control...
Page 22: ...2 Unit description 6 KaVo Dental GmbH 214760 rev 1...
Page 32: ...5 Error messages 16 KaVo Dental GmbH 214760 rev 1...
Page 52: ...9 Recommended exposure times 36 KaVo Dental GmbH 214760 rev 1...