4 Default settings
214760 rev 1
KaVo Dental GmbH
13
4 Default settings
4.1 Program mode
Enter or exit the program by pressing and holding the kV
button for two seconds or more. The beeper will sound
three times. No exposures may be taken while in the
program mode. The READY light will be off.
The menu selections are scrolled with the UP and DOWN
buttons. Enter or exit the selection by pressing the kV
button. Data is edited with the UP button.
The program mode will exit automatically after it remains
idle for 30 seconds. When control returns to the operating
mode, the display will show the updated exposure time.
Menu Selections
Display Contents
Film Speed
Pr1
fast film (Speed F)
SF
fast film (Ekta speed, E)
SE
slow film (Ultra speed, D)
Sd
Cone selection
Pr2
short round
1
short rectangular
2
long round
3
long rectangular
4
short rectangular for sensor size 1
5
short rectangular for sensor size 2
6
Exposure Counter
Pr3
First (0 - 999)
2 digits displayed
Second (1000 - 99000)
3 digits displayed
AEC selection (NOT IN USE)
Pr4
AEC selection ON
AEC
AEC selection OFF
---
Set Factory Defaults, two beeps
Pr5
Speaker adjustment
Pr6
speaker volume (1 = min. 8 = max.)
1-8
DAP Cumulative Dose Area Product
Pr7
to reset cumulative counter, press D
Sensor type selection*
Pr8
intraoral x-ray sensor (default)
GEN
(alternative intraoral sensor)
SIG
Summary of Contents for 214760
Page 1: ...FOCUS Intraoral X ray User Manual ENGLISH 214760 rev 1 0 805 4936...
Page 2: ......
Page 4: ......
Page 6: ...ii KaVo Dental GmbH 214760 rev 1...
Page 21: ...2 Unit description 214760 rev 1 KaVo Dental GmbH 5 Exposure time control...
Page 22: ...2 Unit description 6 KaVo Dental GmbH 214760 rev 1...
Page 32: ...5 Error messages 16 KaVo Dental GmbH 214760 rev 1...
Page 52: ...9 Recommended exposure times 36 KaVo Dental GmbH 214760 rev 1...