9 Recommended exposure times
214760 rev 1
KaVo Dental GmbH
35
9 Recommended exposure times
Recommended exposure times with digital sensors and
phosphor plates.These are also factory pre-set default
values.
Recommended exposure times with film (F-speed)
60kV, 7mA
70kV, 7mA
9” cone
12” cone
9” cone
12” cone
Adult
Child
Adult
Child
Adult
Child
Adult
Child
Bitewing
0,250 0,160 0,500 0,320 0,125 0,080 0,250 0,160
Maxillary incisor
0,200 0,125 0,400 0,250 0,100 0,063 0,200 0,125
Maxillary cuspid
0,250 0,160 0,500 0,320 0,125 0,080 0,250 0,160
Maxillary molar
0,320 0,200 0,500 0,400 0,160 0,100 0,320 0,200
Occlusal
0,250 0,160 0,500 0,320 0,125 0,080 0,250 0,160
Mandibular incisor
0,200 0,125 0,400 0,250 0,100 0,063 0,200 0,125
Mandibular cuspid
0,250 0,160 0,500 0,300 0,125 0,080 0,250 0,160
Mandibular molar
0,250 0,160 0,500 0,320 0,125 0,080 0,250 0,160
9” cone
60kV
70kV
Adult
Child
Adult
Child
Bitewing
0,320 0,200 0,160 0,100
Maxillary incisor
0,250 0,160 0,125 0,080
Maxillary cuspid
0,320 0,200 0,160 0,100
Maxillary molar
0,400 0,250 0,200 0,125
Occlusal
0,320 0,200 0,160 0,100
Mandibular incisor
0,200 0,125 0,100 0,063
Mandibular cuspid
0,250 0,160 0,125 0,080
Mandibular molar
0,250 0,160 0,125 0,080
Summary of Contents for 214760
Page 1: ...FOCUS Intraoral X ray User Manual ENGLISH 214760 rev 1 0 805 4936...
Page 2: ......
Page 4: ......
Page 6: ...ii KaVo Dental GmbH 214760 rev 1...
Page 21: ...2 Unit description 214760 rev 1 KaVo Dental GmbH 5 Exposure time control...
Page 22: ...2 Unit description 6 KaVo Dental GmbH 214760 rev 1...
Page 32: ...5 Error messages 16 KaVo Dental GmbH 214760 rev 1...
Page 52: ...9 Recommended exposure times 36 KaVo Dental GmbH 214760 rev 1...