4 Default settings
14
KaVo Dental GmbH
214760 rev 1
* Different exposure times can be programmed for each
sensor type (similar way as described in 4.1
Programmable anatomical time settings)
4.2 Programmable anatomical time settings
The anatomical time settings (tooth buttons) have been
preprogrammed by the factory but can be changed if
necessary by the user.
First increase or decrease the exposure time with UP and
DOWN buttons. Then press and hold the corresponding
tooth button for two seconds or more. The beeper will beep
two times. The new time setting is now saved into the
memory.
Phosphor Plate
PHO
Preheat boost adjustment
Pr9
(On, if generator revision 1.x)
On
(Disabled, if generator revision 2.x)
DIS
Menu Selections
Display Contents