8 User's statement
214760 rev 1
KaVo Dental GmbH
29
8 User's statement
Instructions for the use of the unit and precautionary
statements are part of the unit
User's Manual.
RADIATION LEAKAGE TECHNIQUE FACTORS
The maximum-rated peak tube potential is 70 kVp with the
maximum rated continuous tube current of 1.5 mA.
BEAM LIMITING DEVICE / TUBE HOUSING ASSEMBLY
COMPATIBILITY
The tube housing assembly THA-I is compatible with the
beam limiting device.
EQUIPMENT STATEMENT FOR TUBE HOUSING
ASSEMBLY
Art.No
Part number
Cones
Program
Mode Pr2
Cone type
ID
0.805.0057
50410-img
Long cone, round
Ø60 mm, SSD 12” (305 mm)
3
0.805.0058
50420-img
Long cone, rectangular
35 x 45 mm, SSD 12” (305 mm)
4
0.805.0059
50540-img
Short cone, round
Ø60 mm, SSD 9” (229 mm)
1
0.805.0060
50550-img
Short cone, rectangular
35 x 45 mm, SSD, SSD 9” (229 mm)
2
0.805.0061
50551-img
Short cone, rectangular, for Germany
28 x 36 mm, SSD, SSD 9” (229 mm)
5
0.805.0062
50720-img
Long cone, full metal, rectangular
35 x 45 mm, SSD 12” (305 mm)
4
0.805.4235
50750-img
Short cone, full metal, rectangular
35 x 45 mm, SSD, SSD 9” (229 mm)
6
Summary of Contents for 214760
Page 1: ...FOCUS Intraoral X ray User Manual ENGLISH 214760 rev 1 0 805 4936...
Page 2: ......
Page 4: ......
Page 6: ...ii KaVo Dental GmbH 214760 rev 1...
Page 21: ...2 Unit description 214760 rev 1 KaVo Dental GmbH 5 Exposure time control...
Page 22: ...2 Unit description 6 KaVo Dental GmbH 214760 rev 1...
Page 32: ...5 Error messages 16 KaVo Dental GmbH 214760 rev 1...
Page 52: ...9 Recommended exposure times 36 KaVo Dental GmbH 214760 rev 1...