7 Technical data
28
KaVo Dental GmbH
214760 rev 1
Recommended separation distances between portable and mobile RF communications
equipment and the unit.
The unit is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the unit can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the unit as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d
= 1.2
P
80 MHz to 800 MHz
d
= 1.2
P
800 MHz to 2.5 GHz
d
= 2.3
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance
d
in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1. At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2. These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Summary of Contents for 214760
Page 1: ...FOCUS Intraoral X ray User Manual ENGLISH 214760 rev 1 0 805 4936...
Page 2: ......
Page 4: ......
Page 6: ...ii KaVo Dental GmbH 214760 rev 1...
Page 21: ...2 Unit description 214760 rev 1 KaVo Dental GmbH 5 Exposure time control...
Page 22: ...2 Unit description 6 KaVo Dental GmbH 214760 rev 1...
Page 32: ...5 Error messages 16 KaVo Dental GmbH 214760 rev 1...
Page 52: ...9 Recommended exposure times 36 KaVo Dental GmbH 214760 rev 1...