IDA-4 Plus
Operators Manual
B-4
Perform a Dual Flow Rate Test [CnD,a,b,c,d,e,f]
This command starts a Dual Flow rate test on the specified channel. The IDA-4 Plus status
all channels screen will prompt the operator to prime the channel and start the test. Once
primed, the operator can press the start key on the front panel or send the [CnD] command
again.
Parameters:
n = channel number
a = control number of device under test
b = technician name
c = set primary flow rate
d = set primary volume
e = set secondary flow rate
f = set secondary volume
Fetch the Current Flow Measured on a Channel [FLOWn]
Returned to host:
[Flow, nnnn.nn,hh:mm:ss:mmm] <CR><LF>
Where:
nnn.nn is the flow rate in ml/hr
hh:mm:ss:mmm is the time since the start of the test, formatted as:
hh Hours
mmMinutes
ss Seconds
mmm
Milli-seconds
Fetch the Current Volume Accumulated on a Channel [VOLn]
Returned to host:
[Vol,
nnnn.nn,hh:mm:ss:mmm]
<CR><LF>
Where:
nnn.nn is the vol in ml
hh:mm:ss:mmm is the time since the start of the test, formatted as:
hh Hours
mm Minutes
ssSeconds
mmm
Milli-seconds
Summary of Contents for Biomedical IDA-4 Plus Operators
Page 1: ...IDA 4 Plus Infusion Pump Analyzer Operators Manual...
Page 2: ......
Page 4: ...IDA 4 Plus Operators Manual ii...
Page 8: ...IDA 4 Plus Operators Manual vi...
Page 21: ...Infusion Pump Analyzer An Overview 2 2 5 Rear Panel Layout Drawing...
Page 22: ...IDA 4 Plus Operators Manual 2 6...
Page 24: ...IDA 4 Plus Operators Manual 3 2 Inlet Hose Connections Drawing...