Infusion Pump Analyzer
Operating Instructions
3
3-27
PCA Graph Screen
The
PCA GRAPH
screen displays the available data in
a bar graph format. The amplitude of each spike
represents the flow rate at which each bolus is
administered. The width of each spike represents the
time duration of that bolus. The gap between successive
spikes will represent the time interval between boluses.
Numerical representation of the total volume delivered
and the volume of the last bolus delivered will be
displayed down the right of the graph.
Status
This switches to the
STATUS ALL CHANNELS
screen to allow the operator to use another channel or
change settings via the
UTILITIES
menu.
View
This returns to the
PCA TEST
screen in either the
Measuring Mode or the End of Test Mode, depending
on which mode this screen was called from.
PCA Bolus Trigger Methods
The
IDA-4 Plus
has two methods of operating with
PCA bolus triggering; automatic and manual. Some
PCA pumps provide a connector to allow the Patient
Demand to be requested. The
IDA-4 Plus
enables this
facility via the
PCA Trigger / Remote Call Interface
.
Automated Bolus Trigger
With pumps that can be triggered electrically the bolus
trigger signals are generated at one-second intervals,
starting before the end of the lockout period.
The programmed period before lockout time expires is
adjustable from the
TEST PARAMETERS SCREEN
.
Manual Bolus Trigger
It is not possible to trigger some types of PCA pumps
directly. Examples are ambulatory devices that have the
trigger mounted on the pump case or pumps that use a
pneumatic trigger. With these pumps the operator will
have to press the
BOLUS / TRIG
prompt on the
PCA
TEST
screen at the same time as pressing the bolus
request button on the pump.
PCA Pump Testing Scenarios
Various scenarios under which a PCA pump could be
tested are considered in the following paragraphs.
These scenarios may not be exhaustive and some
parameters may have to be changed to satisfy the testing
requirements of some pumps.
PCA Activity Only
The basic test scenario would have the PCA pump set
up to supply an initial loading or bolus, then deliver a
series of boluses following a lockout period. The basal
rate will be set to zero.
C h a n n e l 1 P C A G R A P H
T V o l
3 . 0 0
m l
B o l V o l
1 . 0 0
m l
C o u n t 3
00: 02: 00
STATUS
VIEW
Summary of Contents for Biomedical IDA-4 Plus Operators
Page 1: ...IDA 4 Plus Infusion Pump Analyzer Operators Manual...
Page 2: ......
Page 4: ...IDA 4 Plus Operators Manual ii...
Page 8: ...IDA 4 Plus Operators Manual vi...
Page 21: ...Infusion Pump Analyzer An Overview 2 2 5 Rear Panel Layout Drawing...
Page 22: ...IDA 4 Plus Operators Manual 2 6...
Page 24: ...IDA 4 Plus Operators Manual 3 2 Inlet Hose Connections Drawing...