IDA-4 Plus
Operators Manual
3-12
Device Information
This screen prompts the operator for device and test
information for the specific device under test. Data may
be entered into the fields on this screen from the
external keyboard or the optional bar code reader. The
attributes for each field are described below.
When entering data into these fields the last character
entered may be deleted using the backspace key. If
there is data in the field and
ENTER
is pressed then
that data is accepted. If the operator starts typing at the
first position in the field then any existing data will be
wiped from the field and the newly typed data will
replace it. If
ESC
is pressed at any time during editing a
field then the program will back up to the previous
field, menu or screen as appropriate.
Note: If you enter this screen without an external
keyboard connected, use the up arrow button to back
out of the screen.
Control No
Control Number
is a mandatory field up to sixteen
characters in length. If a test does not have a control
number then it cannot be saved. This is usually a
number or code that uniquely identifies the infusion
device being tested. The cursor will be positioned in
this field when the screen is first displayed.
When field is completed, pressing
ENTER
advances to
the Operator field. Pressing
ESC
in this field returns the
operator back to the
CHANNEL SETUP
menu.
Operator
This is a mandatory field, up to sixteen characters in
length. If a test is to be saved it should be possible to
determine who performed the test.
Pressing
ENTER
at the end of this field advances the
program to the Flow Rate field. Pressing
ESC
will
return to the Control No field.
Flow Rate
This is a mandatory field that only accepts numeric
data. It can accept a maximum of six digits including a
decimal point. This is the flow rate that the infusion
device under test has been set to.
Pressing
ENTER
at the end of this field advances the
program to a sub-menu as described in
SUB MENU 1
.
Pressing
ESC
causes the program to step back to the
Operator field.
Vol
This is the volume you intend to deliver. If this volume
is set, then when the volume is reached the instrument
will give two long beeps and display a message.
D e v i c e I n f o r m a t i o n C h a n 1
Control No
Operator
Flow Rate
Manufacturer
Device Type
Serial Num.
Location
P r e s s E N T E R f o r n e x t f i e l d
P r e s s E S C f o r p r e v i o u s f i e l d
Summary of Contents for Biomedical IDA-4 Plus Operators
Page 1: ...IDA 4 Plus Infusion Pump Analyzer Operators Manual...
Page 2: ......
Page 4: ...IDA 4 Plus Operators Manual ii...
Page 8: ...IDA 4 Plus Operators Manual vi...
Page 21: ...Infusion Pump Analyzer An Overview 2 2 5 Rear Panel Layout Drawing...
Page 22: ...IDA 4 Plus Operators Manual 2 6...
Page 24: ...IDA 4 Plus Operators Manual 3 2 Inlet Hose Connections Drawing...