![Fluke Biomedical IDA-4 Plus Operators Operator'S Manual Download Page 15](http://html1.mh-extra.com/html/fluke/biomedical-ida-4-plus-operators/biomedical-ida-4-plus-operators_operators-manual_2309825015.webp)
Infusion Pump Analyzer
Introduction
1
1-3
Specifications
Electrical Specification:
Supply Voltage
90 - 260 VAC
.
Supply Frequency
50 - 60 Hz.
Supply Power
< 30 VA.
Fuse
20 mm 250 V, 1 A (T) (slow blow).
Earth Leakage Current
< 1.0 mA in single fault condition.
Flow Rate Measurement
Technique:
Flow is calculated by measuring a volume over time.
Range:
0.5 - 1000 ml/hr.
Accuracy:
1% of reading
±
1 LSD for flows of 16 - 200 ml/hr for volumes over 20 ml.
Otherwise, 2% of reading
±
1LSD after delivery of 10 ml under laboratory
conditions.
Volume Measurement
Technique:
Volume is measured directly by the transducer in minimum sample sizes of
60 micro-liters.
Range:
0.06 - 9999 ml.
Accuracy:
1% of reading
±
1 LSD for flows of 16 - 200 ml/hr for volumes over 20 ml.
Otherwise, 2% of reading
±
1LSD after delivery of 10 ml under laboratory
conditions.
PCA Bolus Measurement
Technique:
Volume is measured directly by the transducer in minimum bolus volumes
of 0.5 ml. The measurement is made with a continuous rate between 0.0 and
30 ml/hr. The bolus flow rate should be at least four times the basal flow
rate for reliable detection of boluses.
Min Bolus:
0.5 ml.
Accuracy:
See volume measurement.
Pressure Measurement
Technique:
Direct occlusion of the infusion line and measurement of pressure prior to
the glass transducer.
Range:
0 to 45 PSI and equivalents in mmHg and kPa.
Accuracy:
1% of Full Scale
±
1 LSD under laboratory conditions.
Summary of Contents for Biomedical IDA-4 Plus Operators
Page 1: ...IDA 4 Plus Infusion Pump Analyzer Operators Manual...
Page 2: ......
Page 4: ...IDA 4 Plus Operators Manual ii...
Page 8: ...IDA 4 Plus Operators Manual vi...
Page 21: ...Infusion Pump Analyzer An Overview 2 2 5 Rear Panel Layout Drawing...
Page 22: ...IDA 4 Plus Operators Manual 2 6...
Page 24: ...IDA 4 Plus Operators Manual 3 2 Inlet Hose Connections Drawing...