3-34
Dual Rate Graph Screen
The
DUAL FLOW RATE
screen displays the average
flow rate for the primary and for the secondary flows.
Deviation marks as described in the single flow rate
graph are shown for the primary flow rate only.
Text at the right of the graph displays the primary flow
rate, volume delivered at that rate and the duration of
the primary flow. This is followed by the secondary
flow rate and volume delivered at the secondary flow
rate.
Status
This switches to the
STATUS ALL CHANNELS
screen to allow the operator to use another channel or
change settings via the
UTILITIES
menu.
View
This returns to the
DUAL RATE TEST
screen in either
the Measuring Mode or the End of Test Mode,
depending on which mode called this screen.
C H A N N E L 1 D U A L F L O W
Primary
F 55. 25
V 0. 93
00: 01: 00
Secondary
F 5. 19
V 0. 35
00: 04: 01
00: 05: 06