![Fluke Biomedical IDA-4 Plus Operators Operator'S Manual Download Page 52](http://html1.mh-extra.com/html/fluke/biomedical-ida-4-plus-operators/biomedical-ida-4-plus-operators_operators-manual_2309825052.webp)
IDA-4 Plus
Operators Manual
3-30
Dual Rate Test Information Screen
This screen prompts the operator for information
required to perform a dual rate flow test. As a minimum
requirement the primary flow rate will be needed.
Secondary Rate
This is the flow rate of the second part of the infusion.
This is a mandatory field for dual flow rate
measurement. It is used in conjunction with the primary
flow rate to help the flow rate discrimination
algorithms. Secondary will show as Flow 1.
Pressing
ENTER
in this field advances the program to
the Secondary Volume field. Pressing
ESC
will step
back to the Primary Volume field.
Secondary Volume
This is the volume to be delivered at the secondary flow
rate. This value is recorded for information only.
Pressing
ENTER
in this field advances to the
DUAL
RATE TEST
screen. Pressing
ESC
will backtrack to
the Secondary Rate field.
Primary Rate
This is the flow rate of the first part of the infusion. If a
Flow Rate has been entered on the
DEVICE
INFORMATION
screen it is automatically entered
into this field. This is a mandatory field for dual flow
rate measurement. It is used in conjunction with the
secondary flow rate to help the flow rate discrimination
algorithms. Primary will show as Flow 2.
Pressing
ENTER
in this field advances the program to
the Primary Volume field. Pressing
ESC
will return to
the
CHANNEL SETUP
menu screen.
Primary Volume
This is the volume to be delivered at the primary flow
rate. This value is recorded for information only.
Pressing
ENTER
in this field advances the program to
the Secondary Rate field. Pressing
ESC
will step back
to the Primary Rate field.
D U A L R A T E T E S T I N F O R M A T I O N
S e c o n d a r y R a t e
ml
/h
S e c o n d a r y V o l
ml
P r i m a r y R a t e
ml /h
P r i m a r y V o l
ml
0 1 2 3 4 5 6 7 8 9 .
O K
E S C D E L
Summary of Contents for Biomedical IDA-4 Plus Operators
Page 1: ...IDA 4 Plus Infusion Pump Analyzer Operators Manual...
Page 2: ......
Page 4: ...IDA 4 Plus Operators Manual ii...
Page 8: ...IDA 4 Plus Operators Manual vi...
Page 21: ...Infusion Pump Analyzer An Overview 2 2 5 Rear Panel Layout Drawing...
Page 22: ...IDA 4 Plus Operators Manual 2 6...
Page 24: ...IDA 4 Plus Operators Manual 3 2 Inlet Hose Connections Drawing...