Perfomance Testing and Calibration
Performance Tests
4
4-11
HI
HI
LO
LO
HI
OUTPUT
V A
SENSE
V
AUX
CURRENT
GUARD GROUND
WIDEBAND
Ω
Ω
5700A Calibrator
FREEZE
DISPLAY
MODE
SAVE
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
1
2
5
3
4
A
COM
V
EXT TRIG
mA/
A
os16f.eps
Figure 4-4. Configuration 5 (Ohms)
Capacitance Test
4-10.
Rotate the GMM selector to Diode/Capacitance Test (
ä
). Connect the GMM
Ë
and
â
inputs to the GenRad 1412BC. Then perform the test steps in Table 4-9.
Table 4-9. Capacitance Performance Test
Step
GMM Range
Input
Input Source
Lower Limit
Upper Limit
1
10000 pF
open
open
0
130
2
1
µ
F
0.950
µ
F
GenRad 1412-BC
0.930
0.970