867B,863
Service Manual
1-8
Resolution and Accuracy
The following specifications apply within 1 year of calibration when operating in a temperature range of
18°C (64°F) to 28°C (82°F).
AC Volts (True RMS, AC-Coupled) [
±
(percent of r counts)]
Frequency
Range
Res.
20 -
50 Hz
50 -
1 kHz
1 kHz -
30 kHz
30 kHz-
100 kHz
100kHz-
200kHz
200 kHz-
300 kHz
300.00 mV
0.01 mV
1.5% + 10
0.19 dB
0.5% + 10
0.10 dB
0.5% + 10
0.10 dB
4% + 200
1.39 dB
8% + 200
1.68 dB
10%+ 200
1.82 dB
3.0000V
0.1 mV
1.5% + 10
0.19 dB
0.5% + 10
0.10 dB
0.5% + 10
0.10 dB
4% + 200
1.39 dB
8% + 200
1.68 dB
10%+ 200
1.82 dB
30.000V
1 mV
1.5% + 10
0.19 dB
0.5% + 10
0.10 dB
0.5% + 10
0.10 dB
4% + 200
1.39 dB
8% + 200
1.68 dB
10%+ 200
1.82 dB
300.00V
10 mV
1.5% + 10
0.19 dB
0.5% + 10
0.10 dB
0.5% + 10
0.10 dB
4% + 200
1.39 dB
8% + 200
1.68 dB
10%+ 200
1.82 dB
1000.0V
100 mV
1.5% + 10
0.19 dB
1.5% + 10
0.19 dB
NA
NA
NA
NA
300 mV - 300V ranges
≥
3:1, 1000V range
≥
3:1 decreasing to
≥
1.41:1 as input voltage increases to 1000V (peak
voltage not to exceed 1414V).
Measurement Range:
300 mV range from 10% to 100% of range.
3V - 1000V ranges from 5% to 100% of range.
For frequencies > 100 kHz 30% to 100% of range (all ranges).
Summary of Contents for 863
Page 49: ...Maintenance Disassembly 3 3 5 OFF 1 6 Places 4 5 5 3 3 2 os5f eps Figure 3 1 Disassembly ...
Page 50: ...867B 863 Service Manual 3 6 os6f eps Figure 3 1 Disassembly cont ...
Page 52: ...867B 863 Service Manual 3 8 5 1 6 2 3 4 os30f eps Figure 3 2 Reassembly ...
Page 96: ...6 1 Chapter 6 Schematic Diagrams Title Page 6 1 A1 Main PCA Assembly 6 3 ...
Page 98: ...867B 863 Service Manual 6 4 FLUKE 867 1001 Sheet 2 of 5 Figure 6 1 A1 Main PCA cont ...
Page 99: ...Schematic Diagrams 6 6 5 FLUKE 867 1001 Sheet 3 of 5 Figure 6 1 A1 Main PCA cont ...
Page 100: ...867B 863 Service Manual 6 6 FLUKE 867 1001 Sheet 4 of 5 Figure 6 1 A1 Main PCA cont ...
Page 101: ...Schematic Diagrams 6 6 7 FLUKE 867 1001 Sheet 5 of 5 Figure 6 1 A1 Main PCA cont ...