867B,863
Service Manual
4-10
Ohms and Conductance Tests
4-9.
Rotate the GMM selector to Ohms (
ò
). Connect the GMM and the 5700A in
Configuration 5 (Figure 4-4). Then perform the test steps in Table 4-7 (for ohms) and
Table 4-8 (for conductance).
Before proceeding to the next test, place the 5700A in Standby.
Table 4-7. Ohms Performance Test
Step
Range
Input
Input Source
Lower Limit
Upper Limit
1
300.00
Ω
0.00
Ω
Shorting Bar
0.00
0.10
2
300.00
Ω
190.00
Ω
5700A
189.77
190.23
3
3.0000 k
Ω
1.9000 k
Ω
5700A
1.8985
1.9015
4
30.000 k
Ω
19.000 k
Ω
5700A
18.985
19.015
5
300.00 k
Ω
190.00 k
Ω
5700A
189.85
190.15
6
3.0000 M
Ω
1.9000 M
Ω
5700A
1.8970
1.9031
7
30.000 M
Ω
10.000 M
Ω
5700A
9.977
10.023
Table 4-8. Conductance Performance Test
Step
Range
Input
Input Source
Lower Limit
Upper Limit
1
300.00 nS
10.00 M
Ω
5700A
99.30
100.70
2
3000.0 nS
1.0000 M
Ω
5700A
993.0
1007.0
Summary of Contents for 863
Page 49: ...Maintenance Disassembly 3 3 5 OFF 1 6 Places 4 5 5 3 3 2 os5f eps Figure 3 1 Disassembly ...
Page 50: ...867B 863 Service Manual 3 6 os6f eps Figure 3 1 Disassembly cont ...
Page 52: ...867B 863 Service Manual 3 8 5 1 6 2 3 4 os30f eps Figure 3 2 Reassembly ...
Page 96: ...6 1 Chapter 6 Schematic Diagrams Title Page 6 1 A1 Main PCA Assembly 6 3 ...
Page 98: ...867B 863 Service Manual 6 4 FLUKE 867 1001 Sheet 2 of 5 Figure 6 1 A1 Main PCA cont ...
Page 99: ...Schematic Diagrams 6 6 5 FLUKE 867 1001 Sheet 3 of 5 Figure 6 1 A1 Main PCA cont ...
Page 100: ...867B 863 Service Manual 6 6 FLUKE 867 1001 Sheet 4 of 5 Figure 6 1 A1 Main PCA cont ...
Page 101: ...Schematic Diagrams 6 6 7 FLUKE 867 1001 Sheet 5 of 5 Figure 6 1 A1 Main PCA cont ...