867B,863
Service Manual
4-6
DC Volts Test
4-6.
Connect the GMM and 5700A in Configuration 2 (Figure 4-2). Then perform the three
steps called for in Table 4-3. Before proceeding to the next test, place the 5700A in
Standby.
Table 4-3. DC Volts Performance Test
Model 867B
Model 863
Step
GMM
Range
Input
Input Source
Lower
Limit
Upper
Limit
Lower
Limit
Upper
Limit
1
30.000V
29.000V
5700A
28.991
29.009
28.986
29.014
2
300.00V
-290.00V
5700A
-290.09
-289.91
-290.14
-289.86
3
1000.0V
1000.0V
5700A
999.5
1000.4
999.4
1000.6
HI
HI
LO
LO
HI
OUTPUT
V A
SENSE
V
AUX
CURRENT
GUARD GROUND
WIDEBAND
Ω
Ω
5700A Calibrator
FREEZE
DISPLAY
MODE
SAVE
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
1
2
5
3
4
A
COM
V
EXT TRIG
mA/
A
V
os13f.eps
Figure 4-2. Configuration 2 (DC Volts)
Summary of Contents for 863
Page 49: ...Maintenance Disassembly 3 3 5 OFF 1 6 Places 4 5 5 3 3 2 os5f eps Figure 3 1 Disassembly ...
Page 50: ...867B 863 Service Manual 3 6 os6f eps Figure 3 1 Disassembly cont ...
Page 52: ...867B 863 Service Manual 3 8 5 1 6 2 3 4 os30f eps Figure 3 2 Reassembly ...
Page 96: ...6 1 Chapter 6 Schematic Diagrams Title Page 6 1 A1 Main PCA Assembly 6 3 ...
Page 98: ...867B 863 Service Manual 6 4 FLUKE 867 1001 Sheet 2 of 5 Figure 6 1 A1 Main PCA cont ...
Page 99: ...Schematic Diagrams 6 6 5 FLUKE 867 1001 Sheet 3 of 5 Figure 6 1 A1 Main PCA cont ...
Page 100: ...867B 863 Service Manual 6 6 FLUKE 867 1001 Sheet 4 of 5 Figure 6 1 A1 Main PCA cont ...
Page 101: ...Schematic Diagrams 6 6 7 FLUKE 867 1001 Sheet 5 of 5 Figure 6 1 A1 Main PCA cont ...