5500A
Service Manual
6-12
6-17. Overload Measurement Specifications
Table 6-14. Overload Measurement Specifications
Source
Voltage
Typical ‘On’ current
indication
Typical ‘Off’ current
indication
Maximum Time Limit DC or
AC (1 kHz)
5 V to 9 V
100 mA to 180 mA
10 mA
setable 1 s to 60 s
6-18. Theory of Operation
The following discussion provides a brief overview of the following SC600 operating
modes: voltage, edge, leveled sine wave, time marker, wave generator, video, pulse
generator, input impedance, and overload. This discussion will allow you to identify
which of the main plug-in boards of the Calibrator Mainframe are defective. Figure 6-1
shows a block diagram of the SC600 Option, also referred to as the A50 board.
Functions that are not depicted in the figure are generated from the DDS Assembly (A6
board). For a diagram of all Calibrator Mainframe board assemblies, refer to Figure 2-1.
6-19. Voltage
Mode
All signals for the voltage function are generated from the A51 Voltage/Video board, a
daughter card to the A50 board. A dc reference voltage is supplied to the A51 board
from the A6 DDS board; all dc and ac oscilloscope output voltages are derived from this
signal and generated on the A51 board. The output of the A51 board is passed to the A50
Signal board (also attached to the A50 board) and attenuator module and is then cabled
to the output connectors on the front panel. The reference dc signal is used to generate
both + and - dc and ac signals that are amplified or attenuated to provide the complete
range of output signals.
6-20. Edge
Mode
The edge clock originates on the DDS A6 board and is passed to the A50 board. The
signal is then shaped and split to generate the fast edge and external trigger signals. The
edge signal is passed from the A50 board first to the attenuator assembly (where range
attenuation occurs) and then to the SCOPE connector BNC on the front panel. If turned
on, the trigger is connected to the Trig Out BNC on the front panel.
6-21. Leveled Sine Wave Mode
All of the leveled sine wave signals (from 50 kHz to 600 MHz) are produced on the A50
board. The leveled sine wave signal is passed from the A50 board to the on-board
attenuator assembly. The attenuator assembly provides range attenuation and also
contains a power detector which maintains amplitude flatness across the frequency
range. The signal is then passed to the SCOPE connector BNC on the front panel.
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......