SC300 Option
SC300 Hardware Adjustments for the A4 Board
6
6-115
6-144. SC300 Hardware Adjustments for the A4 Board
Hardware adjustments must be made to the leveled sine and edge functions each time the
SC300 is repaired. In addition to the adjustment procedures, this section provides lists of
the required equipment and some recommendations on models that have the capabilities
required by these procedures. Equivalent models can be substituted if necessary.
6-145. Equipment Required
The following equipment is necessary for performing the hardware adjustments
described in this section. The models listed are recommended for providing accurate
results.
•
Standard adjustment tool for adjusting the pots and trimmer caps
•
Extender Card
•
Oscilloscope Mainframe and Sampling Head (Tektronix 11801B with SD-22)
•
Delay Cable, 60 ns
•
Spectrum Analyzer (Hewlett Packard 8590A)
6-146. Adjusting the Leveled Sine Wave Function
There are two adjustment procedures that need to be made for the leveled sine wave
function. The first procedure adjusts the balance out of the LO VCO so that the signal is
balanced between the two VCOs. The second procedure adjusts the harmonics.
6-147. Equipment Setup
This procedure uses the spectrum analyzer. Before you begin this procedure, verify that
the Calibrator Mainframe is in leveled sine wave mode (the Levsine menu is displayed),
and program it to output 5.5V p-p @ 110 MHz. Press
O
to activate the output.
Connect the Calibrator Mainframe to the Spectrum Analyzer. Adjust the Spectrum
Analyzer so that it displays one peak across its horizontal center line. The far right of the
peak is fixed at the far right of the center line, as shown below.
6-148. Adjusting the Leveled Sine Wave VCO Balance
Once you have completed the setup described above, perform the following procedure to
adjust the VCO balance for the leveled sine wave function.
1.
Program the Calibrator Mainframe for an output of 5.5V @ 110 MHz.
2.
Set the Spectrum Analyzer to the parameters listed below.
Spectrum Analyzer Setup
Start Frequency
110 MHz
Stop Frequency
113 MHz
Resolution Bandwidth
30 kHz
Video Bandwidth
3 kHz
Reference Level
20 dBm
The Spectrum Analyzer will display a spur in the waveform approximately 1 MHz
away from the carrier frequency. Refer to Figure 6-31 to identify the spur.
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......