SC600 Option
Verification
6
6-57
6-70. Pulse Period Verification
This procedure uses the following equipment:
•
PM 6680 Frequency Counter with an ovenized timebase (Option PM 9690 or PM
9691)
•
BNC cable supplied with the SC600
Refer to Figure 6-6 for the proper setup connections. Set the Calibrator Mainframe to
SCOPE mode, with the Pulse menu on the display. Press
O
on the Calibrator
Mainframe to activate the output. Then follow these steps to verify the Pulse period.
1.
Set the PM 6680’s FUNCTION to measure period on channel A with auto trigger,
measurement time set to 1 second or longer, 50
Ω
impedance, and filter off.
2.
Using the BNC cable, connect the SCOPE connector on the Calibrator Mainframe to
PM 6680 channel A.
3.
Program the Calibrator Mainframe to output the pulse width and period (at 2.5 V) as
listed in Table 6-44.
4.
Allow the PM 6680 reading to stabilize, then record the PM 6680 reading for each
period listed for the Calibrator Mainframe. Compare to the tolerance column of
Table 6-44.
Table 6-44. Pulse Period Verification
Calibrator Mainframe
Output
PM 6680 Reading
Width
Period
(Period)
Tolerance
80 ns
200 ns
5E-13 s
500 ns
10 ms
2.5E-08 s
500 ns
20 ms
5.0E-08 s
6-71. MeasZ Resistance Verification
The MeasZ resistance function is verified by measuring resistors of known values. The
measurement value is then compared to the resistor actual value.
The resistors must make a solid connection to a BNC(f) to enable a connection to the end
of the BNC cable supplied with the SC600. The resistance values must be known at this
BNC(f) connector. Fluke uses an HP 3458A DMM to make a 4-wire ohms measurement
at the BNC(f) connector to determine the actual resistance values.
This procedure uses the following equipment:
•
Resistors of known values: 1.5 M
Ω
, 1 M
Ω
, 60
Ω
, 50
Ω
, 40
Ω
nominal
•
adapters to connect resistors to BNC(f) connector
•
BNC cable supplied with the SC600
Refer to Figure 6-17 for the proper setup connections.
Set the Calibrator Mainframe to SCOPE mode, with the MeasZ menu on the display.
Then follow these steps to verify the MeasZ resistance function.
1.
Set the Calibrator Mainframe MeasZ resistance range as indicated in Table 6-45.
(The blue softkey under MEASURE toggles the MeasZ ranges).
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......