SC600 Option
SC600 Hardware Adjustments
6
6-63
Note
To verify the edge aberrations back to national standards, you should send
your Calibrator Mainframe to Fluke, or other facility that has established
traceability for aberrations. Fluke, for example, has a reference pulse that
is sent to the National Institute of Standards and Technology (NIST) for
characterization. This information is then transferred to high speed
sampling heads, which are used to adjust and verify the SC600.
6-81. Equipment Setup
The following equipment is needed for this procedure:
•
Oscilloscope: Tektronix 11801 with SD22/26 input module or Tektronix TDS 820
with 8 GHz bandwidth.
•
10 dB Attenuator: Weinschel 9-10 (SMA) or Weinschel 18W-10 or equivalent
•
Output cable provided with the SC600
Before you begin this procedure, verify that the SC600 is in the edge mode (the Edge
menu is displayed), and program it to output 1 V p-p @ 1 MHz. Press
O
to activate
the output.
Refer to Figure 6-7 for the proper setup connections and connect the Calibrator
Mainframe to the oscilloscope. Set the oscilloscope vertical to 10 mV/div and horizontal
to 1 ns/div. Set the oscilloscope to look at the 90% point of the edge signal; use this
point as the reference level. Set the oscilloscope to look at the first 10 ns of the edge
signal with the rising edge at the left edge of the oscilloscope display.
6-82. Adjusting the Edge Aberrations
Refer to Figure 6-17 while making the following adjustments:
1.
Adjust A90R13 to set the edge signal at the right edge of oscilloscope display, at 10
ns, to the reference level set above.
2.
Adjust A90R36 so the first overshoot is the same amplitude as the next highest
aberration.
3.
Adjust A90R35 so that the second and third overshoot aberrations are the same
amplitude as the first aberration.
4.
Adjust A90R12 to set the edge signal occurring between 2 ns and 10 ns to the
reference level set above.
5.
Readjust A90R36 and A90R35 to obtain equal amplitudes for the first, second, and
third aberrations.
6.
Adjust A90R13 to set the edge signal occurring between 0 ns and 2 ns to the
reference point set above. Center any aberrations so the peaks are equal above and
below the reference level.
7.
Readjust A90R12 if necessary to keep the edge signal occurring between 2 ns and 10
ns at the reference level.
8.
Readjust A90R13 if necessary to keep the edge signal occurring between 0 ns and 2
ns at the reference level.
9.
Set the UUT output to 250 mV and the oscilloscope vertical to 2 mV/div. Check the
aberrations.
10.
Connect the 10 dB attenuator to the oscilloscope input. Connect the UUT to the
attenuator and program the UUT output to 2.5 V.
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......