Introduction and Specifications
Specifications
1
1-9
1-7. Resistance
Specifications
Ranges
[1]
Absolute Uncertainty, tcal ± 5
°
C
± (% of
Ω
) [2]
Resolution
Allowable
Current
90 days
1 year
[4]
0 to 10.99
Ω
0.009%
0.008
Ω
[3]
0.012%
0.008
Ω
[3]
0.001
Ω
1 to 125 mA
11 to 32.999
Ω
0.009
0.015 [3]
0.012
0.015 [3]
0.001
1 to 125 mA
33 to 109.999
Ω
0.007
0.015 [3]
0.009
0.015 [3]
0.001
1 to 70 mA
110 to 329.999
Ω
0.007
0.015 [3]
0.009
0.015 [3]
0.001
1 to 40 mA
330
Ω
to 1.09999 k
Ω
0.007
0.06
0.009
0.06
0.01
250
µ
A to 18 mA
1.1 to 3.29999 k
Ω
0.007
0.06
0.009
0.06
0.01
250
µ
A to 5 mA
3.3 to 10.9999 k
Ω
0.007
0.6
0.009
0.6
0.1
25
µ
A to 1.8 mA
11 to 32.9999 k
Ω
0.007
0.6
0.009
0.6
0.1
25
µ
A to 0.5 mA
33 to 109.999 k
Ω
0.008
6
0.011
6
1
2.5
µ
A to 0.18 mA
110 to 329.999 k
Ω
0.009
6
0.012
6
1
2.5
µ
A to 0.05 mA
330k to 1.09999 M
Ω
0.011
55
0.015
55
10
250 nA to 0.018 mA
1.1 to 3.29999 M
Ω
0.011
55
0.015
55
10
250 nA to 5
µ
A
3.3 to 10.9999 M
Ω
0.045
550
0.06
550
100
25 nA to 1.8
µ
A
11 to 32.9999 M
Ω
0.075
550
0.1
550
100
25 nA to 0.5
µ
A
33 to 109.999 M
Ω
0.4
5.5k
0.5
5.5k
1000
2.5 nA to 0.18
µ
A
110 to 330 M
Ω
0.4
16.5k
0.5
16.5k
1000
2.5 nA to 0.06
µ
A
[1]
Continuously variable from 0 to 330 M
Ω
.
[2]
Applies for COMP OFF (to the 5500A Calibrator front panel NORMAL terminals) and 2-wire and
4-wire compensation.
[3]
The floor adder is improved to 0.006
Ω
(0 to 10.99
Ω
range) and 0.010
Ω
(11 to 329.999
Ω
) if the
5500A Calibrator is zeroed (ohms zero or instrument zero) within 8 hours and temperature is ±1
°
C
of zeroing ambient temperature.
[4]
Do not exceed the largest current for each range. For currents lower than shown, the floor adder
increases by Floor
(new)
= Floor
(old)
x I
min
/I
actual
. For example, a 100
µ
A stimulus measuring 100
Ω
has a
floor uncertainty of 0.01
Ω
x 1 mA/100
µ
A = 0.1
Ω
.
Ranges
Maximum Voltage [1]
Maximum Lead Resistance [2]
0 to 10.99
Ω
1.37 V
<3.2
Ω
11 to 32.999
Ω
4.12
<3.2
33 to 109.999
Ω
7.7
<3.2
110 to 329.999
Ω
13.2
<3.2
330
Ω
to 1.09999 k
Ω
19.8
<6
1.1 to 3.29999 k
Ω
16.5
<6
3.3 to 10.9999 k
Ω
19.8
<6
11 to 32.9999 k
Ω
16.5
<6
33 to 109.999 k
Ω
19.8
<6
110 to 329.999 k
Ω
16.5
(n/a 110 k
Ω
and above)
330 k
Ω
to 1.09999 M
Ω
19.8
“
1.1 to 3.29999 M
Ω
16.5
“
3.3 to 10.9999 M
Ω
19.8
“
11 to 32.9999 M
Ω
16.5
“
33 to 109.999 M
Ω
19.8
“
110 to 330 M
Ω
19.8
“
[1] This is for the largest resistance for each range. The maximum voltage for other values is Imax
(highest value of Allowable Current above) multiplied by Rout.
[2] Maximum lead resistance for no additional error in 2-wire COMP.
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......