SC600 Option
Verification
6
6-33
6-49. Verification at 50
Ω
For the 50
Ω
verification, connect the Calibrator Mainframe’s SCOPE connector to the
HP 3458A input, using the cable supplied with the Calibrator Mainframe, the external
50
Ω
termination, and the BNC(f) to Double Banana adapter. (The 50
Ω
termination is
closest to the HP 3458A input.) Connect the Calibrator Mainframe TRIG OUT
connector to the HP 3458A Ext Trig connector located on the rear of that instrument.
Make sure the Calibrator Mainframe impedance is set to 50
Ω
. (The blue softkey under
Output @ toggles the impedance between 50
Ω
and 1 M
Ω
). Proceed with the following
steps:
1.
Set the HP 3458A to DCV, NPLC = .01, TRIG EXT, and the DELAY to .0007 for
measuring the topline of the wave form, and the DELAY to .0012 for measuring the
baseline of the wave form. Manually lock the HP 3458A to the range that gives the
most resolution for the topline measurements. Use this same range for the
corresponding baseline measurements at each step. See Table 6-22.
2.
Enable the Calibrator Mainframe external trigger by toggling the blue softkey under
TRIG to /1.
3.
Measure the topline first, as indicated in Table 6-22. For each measurement, take
samples for at least two seconds, using the HP 3458A MATH functions to determine
the average or mean value. See “Setup for SC600 Voltage Square Wave
Measurements” for more details.
4.
Measure the baseline of each output after the corresponding topline measurement, as
indicated in Table 6-22. The peak-to-peak value is the difference between the topline
and baseline measurements. Compare the result to the tolerance column.
Table 6-22. AC Voltage Verification at 50
Ω
Calibrator
Mainframe
Output
(1 kHz)
HP 3458A
Range
Topline
Reading
Baseline
Reading
Peak-to-Peak
Peak-to-Peak x
Correction
Tolerance
(
±
V)
1 mV
100 mV dc
0.000043
-1 mV
100 mV dc
0.000043
10 mV
100 mV dc
0.000065
-10 mV
100 mV dc
0.000065
25 mV
100 mV dc
0.000103
-25 mV
100 mV dc
0.000103
110 mV
100 mV dc
0.000315
-110 mV
100 mV dc
0.000315
500 mV
1 V dc
0.00129
-500 mV
1 V dc
0.00129
2.2 V
10 V dc
0.00554
-2.2 V
10 V dc
0.00554
6.6 V
10 V dc
0.01654
-6.6 V
10 V dc
0.01654
Summary of Contents for 5500A
Page 12: ...5500A Service Manual x...
Page 16: ...5500A Service Manual 1 2...
Page 46: ...5500A Service Manual 1 32...
Page 48: ...5500A Service Manual 2 2...
Page 98: ...5500A Service Manual 4 2...
Page 102: ...5500A Service Manual 4 6 om017f eps Figure 4 2 Exploded View of Front Panel Assemblies...
Page 116: ...5500A Service Manual 5 2...
Page 128: ...5500A Service Manual 6 2...
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......
Page 266: ......
Page 267: ......
Page 268: ......
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 275: ......
Page 276: ......
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ...DS2E SL2 DC5V RELAY...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......