Remote Commands
Remote Command Listing
5
5-19
TC_TYPE?
Sequential
Overlapped
IEEE-488
RS-232
x
x
x
x
(Thermocouple Type query) Returns the Thermocouple (TC) temperature sensor
type. When the thermocouple type is changed while simulating a temperature
output, the temperature is changed to 0
°
C.
Responses:
B
(B-type thermocouple)
C
(C-type thermocouple)
E
(E-type thermocouple)
J
(J-type thermocouple)
K
(K-type thermocouple)
L
(L-type thermocouple)
N
(N-type thermocouple)
R
(R-type thermocouple)
S
(S-type thermocouple)
T
(T-type thermocouple)
Z
(1 mV/
°
C linear output)
Example:
TC_TYPE?
returns
K
Returns
K
when the thermocouple type for simulating a temperature output is a
K-type thermocouple.
TSENS_TYPE
Overlapped
IEEE-488
RS-232
Sequential
x
x
x
x
(Temperature Sensor Type command) Sets the temperature sensor type to
thermocouple (TC) or Resistance Temperature Detector (RTD) for temperature
measurements.
Parameters:
TC
(Thermocouple)
RTD
(Resistance Temperature Detector)
Example:
TSENS_TYPE RTD
Set the temperature sensor type to an RTD.