525A
Users Manual
5-12
PRES_UNIT?
Sequential
Overlapped
IEEE-488
RS-232
x
x
x
x
(Pressure Units query) Returns the pressure display units.
Responses:
PSI
(pounds per square inch)
INH2O4C
(inches of water at 4 degrees Celsius)
INH2O20C
(inches of water at 20 degrees Celsius)
CMH2O4C
(centimeters of water at 4 degrees Celsius)
CMH2O20C
(centimeters of water at 20 degrees Celsius)
BAR
(bars)
MBAR
(millibars)
KPAL
(kilopascals)
INHG
(inches of mercury)
MMHG
(millimeters of mercury)
KG/CM2
(kilograms per square centimeter)
Example:
PRES_UNIT?
returns
BAR
RANGE?
Sequential
Overlapped
IEEE-488
RS-232
x
x
x
x
(Range query) Returns the present voltage output range.
Response:
<range>
Examples:
V_0.1V
(dc volts 100 mV range)
V_1V
(dc volts 1V range)
V_10V
(dc volts 10 V range)
V_100V
(dc volts 100 V range)
Returns the symbolic name of the voltage output range.
RANGELCK
Sequential
Overlapped
IEEE-488
RS-232
x
x
x
x
(Range lock command) Locks or unlocks the present voltage range.
Parameter:
ON
Locks the present voltage range
OFF
Unlocks the present voltage range.