Remote Commands
Remote Command Listing
5
5-15
RTD_TYPE?
Sequential
Overlapped
IEEE-488
RS-232
x
x
x
x
(Resistance Temperature Detector Type query) Returns the Resistance
Temperature Detector (RTD) type used for RTD temperature simulations.
Responses:
PT385_100
(100-ohm RTD, curve
α
=0.00385 ohms/ohm/
°
C)
PT385_200
(200-ohm RTD, curve
α
=0.00385 ohms/ohm/
°
C)
PT385_500
(500-ohm RTD, curve
α
=0.00385 ohms/ohm/
°
C)
PT385_1000
(1000-ohm RTD, curve
α
=0.0038 ohms/ohm/
°
C)
PT392_100
(100-ohm RTD, curve
α
=0.003926 ohms/ohm/
°
C)
PTJIS_100
(100-ohm RTD, curve
α
=0.003916 ohms/ohm/
°
C)
CU10
(10-ohm RTD, empirical curve)
NI120
(120-ohm RTD, empirical curve)
YSI_400
(YSI thermistor curve)
OHMS_HIGH
(4000 ohms range)
OHMS_LOW
(400 ohms range)
SPRT_25
(25 ohm standard PRT with user defined error
coefficients)
USR_DEF<x>
(RTD with user defined coefficients (x from
1 to 5))
Example:
RTD_TYPE?
returns
PTJIS_100
Returns
PTJIS_100
when a 100-ohm RTD with curve
α
=0.3926 ohm/
°
C is set
as the RTD type.
*SRE
Sequential
Overlapped
IEEE-488
RS-232
x
x
x
x
(Service Request Enable command) Loads a byte into the Service Request Enable
(SRE) register. (See “Service Request Enable Register (SRE)” in Chapter 4. Since
bit 6 is not used (decimal value 64), the maximum entry is 255
−
64 = 191.
Parameter:
<value>
(the decimal equivalent of the SRE byte, 0 to 191)
Example:
*SRE 48
Enable bits 4 (MAV), and 5 (ESR).