![Fluke 525A User Manual Download Page 82](http://html.mh-extra.com/html/fluke/525a/525a_user-manual_2309641082.webp)
525A
Users Manual
5-10
OUT
Overlapped
IEEE-488
RS-232
Sequential
x
x
x
x
(Output command) Sets the output of the Calibrator and establishes a new
reference point for the error mode. To source or measure a temperature, select the
desired sensor and sensor parameters first. (See the
TSENS_TYPE
,
RTD_TYPE
,
and
TC_TYPE
commands.)
Use multipliers, k, m, u with the
OUT
command, as desired.
Parameters:
<value>
V
Volts dc
<value>
A
Current dc
<value>
OHM
Resistance
<value>
CEL
Temperature (Celsius)
<value>
FAR
Temperature (Fahrenheit)
Examples:
OUT 15.2 V
(volts; 15.2 V)
OUT 1.2 mA
(current; 1.2 mA)
OUT 5 Ohm
(ohms; 5
Ω
)
OUT 100 CEL
(temperature; 100
°
C)
OUT
−
32 FAR
(temperature;
−
32
°
F)
Each example shows a value and unit, e.g.,
−
15.2 V. If a value is entered without a
unit, the value of the existing output is changed, when logically allowed.
OUT?
Sequential
Overlapped
IEEE-488
RS-232
x
x
x
x
(Output query) Returns the output amplitude of the Calibrator. Multipliers (e.g., K
or M) are not used in the response.
Response:
<amplitude value>,<units>
Examples:
OUT?
returns
−
1.01,V
OUT?
returns
1.88300E
−
02,A
OUT?
returns
1.00,V
OUT?
returns
4.03,OHM
OUT?
returns
1.01,V
OUT?
returns
1.0430E+02,CEL
The respective values for the above examples are:
−
15.2 V
18.83 mA
1.23 V
4 K
Ω
15.2 V
104.3
°
C
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com